qCMOS Camera C15550-20UP / C15550-20UP01 Instruction manual_Ver.1.3
42
(4) Global reset Level trigger mode
Global reset function enables to reset the electric charge of all pixels at the same time. Then all
pixels can start exposure at the same time.
The example below is for the trigger level High. With this Global reset Level trigger mode, the
exposure of all pixels begins when the trigger signal becomes High.
Figure 11-8 (Ex. level High)
* Delay + Jitter = 4H + Maximum. 1H
Standard scan: 1H = 7.2 µs
Ultra quiet scan: 1H = 172.8 µs
Summary of Contents for C15550-20UP
Page 2: ...Blank page...