ONDEMAND* Device Refill Chart
30 min. r
efill
60 min. r
efill
30
5.0
2.5
Refill Time (minutes)
Bolus V
ol
. (ml)
60
The ONDEMAND* device is available in 30 or 60 minute refill times
as labeled on the device. Refill time is approximately linear.
During the Infusion
• A change in appearance and size of the pump may not be
evident during the first 24 hours after start of infusion.
• As medication is delivered, the pump will gradually
become smaller.
• Make sure:
• Clamp is open.
• There are no kinks in the tubing.
• Filter vent is not taped or covered.
• Heat, ice or cold therapy is placed away from the flow
controller.
End of Infusion
• Infusion is complete when pump is no longer inflated.
• Close clamp, disconnect and dispose of pump according to
your institution’s protocol.
Note:
If pump did not perform as expected do not discard.
Contact Halyard Health for product return instructions at:
1-800-448-3569
.
Technical Specifications
Delivery Accuracy:
When filled to the labeled volume,
basal flow rate accuracy is ±15% and bolus dose accuracy
is +10/-20% of the labeled rates when infusion is started
0–8 hours after fill and delivering normal saline as the
diluent at 22°C/72°F.
Typical Flow Curve
The flow rate may be higher or lower at the beginning and
end of the infusion (Figure 4).
Note:
Latex is not in fluid pathway or in contact with human. Refer to
Latex Sensitivity Technical Bulletin at halyardhealth.com.
Storage Conditions
Store under general warehouse conditions. Protect from light
sources and heat. Keep dry.
Figure 4
Typical Flow Curve
Flo
w R
at
e
Percent Delivery Time
100%
Nominal