Halyard ON-Q ONDEMAND Instructions For Use Manual Download Page 8

6

ONDEMAND* Device Refill Chart

30 min. r

efill

60 min. r

efill

30

5.0

2.5

Refill Time (minutes)

Bolus V

ol

. (ml)

60

The ONDEMAND* device is available in 30 or 60 minute refill times 

as labeled on the device. Refill time is approximately linear.

During the Infusion

•   A change in appearance and size of the pump may not be 

evident during the first 24 hours after start of infusion.

•  As medication is delivered, the pump will gradually 

become smaller.

•  Make sure: 

  •   Clamp is open.

  •   There are no kinks in the tubing.

  •   Filter vent is not taped or covered.

•   Heat, ice or cold therapy is placed away from the flow   

 controller.

End of Infusion 

•  Infusion is complete when pump is no longer inflated.  

•  Close clamp, disconnect and dispose of pump according to 

your institution’s protocol.

Note:

 

If pump did not perform as expected do not discard. 

Contact Halyard Health for product return instructions at:  

[email protected] or  

1-800-448-3569

.

Technical Specifications

Delivery Accuracy:

 When filled to the labeled volume,  

basal flow rate accuracy is ±15% and bolus dose accuracy  

is +10/-20% of the labeled rates when infusion is started 

0–8 hours after fill and delivering normal saline as the 

diluent at 22°C/72°F.

Typical Flow Curve 

The flow rate may be higher or  lower at the beginning and 

end of the infusion (Figure 4).

Note:

Latex is not in fluid pathway or in contact with human. Refer to 

Latex Sensitivity Technical Bulletin at halyardhealth.com.

Storage Conditions

Store under general warehouse conditions. Protect from light 

sources and heat. Keep dry.

Figure 4

Typical Flow Curve

Flo

w R

at

e

Percent Delivery Time

100%

Nominal

Summary of Contents for ON-Q ONDEMAND

Page 1: ...Instructions for Use Pain Relief System PUMP with ONDEMAND BOLUS BUTTON...

Page 2: ......

Page 3: ...s no alarm or alert when flow interruption occurs therefore life supporting medications whose usage may cause serious injury or death due to stoppage or under delivery are not recommended for infusion...

Page 4: ...gnificantly increase the flow rate Do not exceed maximum fill volume Table 1 Clamp is provided to stop the infusion Do not remove or break clamp Do not use clamp as an intermittent delivery device Rol...

Page 5: ...dministration include intraoperative site perineural percutaneous and epidural ON Q Pump is indicated to significantly decrease pain and narcotic use when used to deliver local anesthetics to or aroun...

Page 6: ...ion set is primed 4 Replace tubing cap until ready for use 5 Remove the red tab by pulling straight out Figure 3 A It is important to remove red tab completely and ensure it does not break Figure 3 B...

Page 7: ...ce will begin to refill 3 The orange indicator shows how much medication is in the bolus device Figure 3 C 4 The next full bolus will be available when orange indicator is at the top level 5 Pressing...

Page 8: ...ccording to your institution s protocol Note If pump did not perform as expected do not discard Contact Halyard Health for product return instructions at HalyardIrvineProductComplaint hyh com or 1 800...

Page 9: ...rTrademark of Halyard Health Inc or its affiliates 2015 HYH All rights reserved For more information please call 1 949 923 2400 1 800 448 3569 English only or visit www halyardhealth com for the lates...

Page 10: ...8...

Page 11: ......

Page 12: ...lth com Halyard Health Inc 5405 Windward Parkway Alpharetta GA 30004 USA Halyard Belgium BVBA Leonardo Da Vincilaan 1 1930 Zaventem Belgium Sponsored in Australia by Surgical Specialties Pty Ltd 1 17...

Reviews: