5
Warning: Do not pull the red tab upwards
as breakage could occur (Figure 3-B). If red
tab is not removed or breaks while removing,
continuous delivery will occur. This delivery may
be significantly greater than the Total Flow Rate
(bolus + basal).
WRONG WAY
Figure 3-B
6. The device is now ready to use; however, a complete bolus
dose won’t be available until the labeled refill time has
elapsed. The orange indicator should be at the top level
(Figure 3-C).
Starting Infusion
Use Aseptic Technique
1. Connect tubing to patient’s catheter. Make sure
connection is secure.
2. The patient or clinician should give a bolus as soon as
possible after the infusion has started to ensure the bolus
device is working properly. The bolus button should pop up
within a few minutes and the orange indicator should begin
to move towards the top.
Figure 3-A
Bolus Activation
1. Press down on the ONDEMAND* button until it locks into
place (Figure 3-D).
2. Bolus will be delivered and ONDEMAND* device will begin
to refill.
3. The orange indicator shows how much medication is in
the bolus device (Figure 3-C).
4. The next full bolus will be available when orange
indicator is at the top level.
5. Pressing the bolus button prior to the end of the refill
time will result in a partial bolus dose.
Warning: If the bolus button will not latch, close the
clamp. Otherwise continuous medication delivery may
be occurring. This delivery may be significantly greater
than the Total Flow Rate.
Note:
It is normal that it will not latch within 30 minutes of
pressing the bolus button.
Warning: If the ONDEMAND* button does not pop
back up within 30 minutes, check position of orange
indicator:
1. If orange indicator is in the bottom position, close the
clamp. Continuous medication delivery may be occurring
which can be significantly greater than the Total Flow Rate.
or
2. If orange indicator is in the top position, something may
be impeding the flow. Check for tubing kinks, closed
clamp or patency of connected devices such as catheter
or unvented filter (verify patency) according to your
standard protocol.
Orange
Bolus Refill
Indicator
Top
Bottom
Figure 3-C
Figure 3-D
RIGHT WAY