Halyard ON-Q ONDEMAND Instructions For Use Manual Download Page 5

3

temperature and will decrease approximately 1.4% per 

1°F/0.6°C decrease in temperature. 

•   If refrigerated, allow pump to reach room temperature 

before using. 

•  It may take approximately 12 or 15 hours (depending 

on model) for a pump to reach room temperature. (See  

table below.)

Fill Volume (ml)

270

400

Refrigerator to Room Temp (hr)

12

15

Storage

•   Storage of a filled ON-Q* Pump for more than 8 hours prior 

to starting infusion may result in a slower  

flow rate. 

External pressure

•   External pressure such as squeezing or laying on pump 

increases flow rate.

Indications for Use

•  ON-Q* Pump is intended to provide continuous delivery 

of medication (such as local anesthetics) to or around 

surgical wound sites and/or close proximity to nerves 

for preoperative, perioperative and postoperative 

regional anesthesia and/or pain management.  Routes 

of administration include: intraoperative site, perineural, 

percutaneous and epidural.

•  ON-Q* Pump is indicated to significantly decrease pain and 

narcotic use when used to deliver local anesthetics to or 

around surgical wound sites, or close proximity to nerves, 

when compared to narcotic only pain management.

Contraindications

 

•  ON-Q* Pump is not intended for blood, blood products, 

lipids, fat emulsions, or Total Parenteral Nutrition (TPN). 

•    ON-Q* Pump is not intended for intravascular delivery. 

•   The ON-Q* Pump with ONDEMAND* device is 

not intended for use with ON-Q* SOAKER* and 

SILVERSOAKER* catheters because they could prevent 

the proper functioning of the ONDEMAND* device.

Description of Device:  (Figure 1)

The ON-Q* Pump with ONDEMAND* device incorporates a bolus 

device. The ONDEMAND* device delivers a continuous infusion 

(basal) and allows fixed boluses to be delivered on demand by 

patient or healthcare provider.

1

  Fill Port Cap

2

  Fill Port

3

  ON-Q* Pump

4

 Clamp

5

  Air Eliminating Filter

6

   ONDEMAND* Bolus Device

7

 Tubing

6

1

2

3

4

5

Figure 1

5

7

Summary of Contents for ON-Q ONDEMAND

Page 1: ...Instructions for Use Pain Relief System PUMP with ONDEMAND BOLUS BUTTON...

Page 2: ......

Page 3: ...s no alarm or alert when flow interruption occurs therefore life supporting medications whose usage may cause serious injury or death due to stoppage or under delivery are not recommended for infusion...

Page 4: ...gnificantly increase the flow rate Do not exceed maximum fill volume Table 1 Clamp is provided to stop the infusion Do not remove or break clamp Do not use clamp as an intermittent delivery device Rol...

Page 5: ...dministration include intraoperative site perineural percutaneous and epidural ON Q Pump is indicated to significantly decrease pain and narcotic use when used to deliver local anesthetics to or aroun...

Page 6: ...ion set is primed 4 Replace tubing cap until ready for use 5 Remove the red tab by pulling straight out Figure 3 A It is important to remove red tab completely and ensure it does not break Figure 3 B...

Page 7: ...ce will begin to refill 3 The orange indicator shows how much medication is in the bolus device Figure 3 C 4 The next full bolus will be available when orange indicator is at the top level 5 Pressing...

Page 8: ...ccording to your institution s protocol Note If pump did not perform as expected do not discard Contact Halyard Health for product return instructions at HalyardIrvineProductComplaint hyh com or 1 800...

Page 9: ...rTrademark of Halyard Health Inc or its affiliates 2015 HYH All rights reserved For more information please call 1 949 923 2400 1 800 448 3569 English only or visit www halyardhealth com for the lates...

Page 10: ...8...

Page 11: ......

Page 12: ...lth com Halyard Health Inc 5405 Windward Parkway Alpharetta GA 30004 USA Halyard Belgium BVBA Leonardo Da Vincilaan 1 1930 Zaventem Belgium Sponsored in Australia by Surgical Specialties Pty Ltd 1 17...

Reviews: