2
If the bolus button does not pop back up within
30 minutes of pressing it, check position of orange
indicator:
• If orange indicator is in the bottom position, close
the clamp. Continuous medication delivery may
be occurring significantly greater than the Total
Flow Rate.
• If orange indicator is in the top position:
something may be impeding the flow. Check
for tubing kinks, closed clamp or patency of
connected devices such as catheter or unvented
filter (verify patency) according to your standard
protocol.
• It is the responsibility of the healthcare provider to
ensure patient is educated on the proper use of the
system.
• It is the responsibility of the healthcare provider to
modify Patient Guidelines provided with the pump
as appropriate for your patients’ clinical status and
medication prescribed.
Cautions
•
Do not use if package is open, damaged or a
protector cap is missing.
•
Single use only. Do not resterilize, refill or reuse.
Reuse of the device could result in the following risks:
• Improper functioning of the device (i.e., inaccurate flow
rate)
• Increased risk of infection
• Occlusion of the device (i.e., impedes or stops infusion)
• The pump is sterile and non-pyrogenic.
•
Product uses Di (2-ethylhexyl) phthalate (DEHP)
plasticized PVC:
• DEHP is a commonly used plasticizer in medical devices.
There is no conclusive scientific evidence to date that
exposure to DEHP has a harmful effect on humans.
However, the risk and benefit of using medical devices
with DEHP for pregnant women, breastfeeding mothers,
infants and children should be evaluated prior to use.
• Certain solutions may be incompatible with the
PVC material used in the administration set. Consult
drug package insert and other available sources of
information for a more thorough understanding of
possible incompatibility problems.
• Do not underfill pump. Underfilling pump may significantly
increase the flow rate.
• Do not exceed maximum fill volume. (Table 1)
• Clamp is provided to stop the infusion. Do not remove or
break clamp. Do not use clamp as an intermittent
delivery device.
• Roll tubing between fingers to promote flow if clamped for
extended time.
• The fill volume, infusion rate, bolus dose and bolus interval
is labeled on the fill port.
• Avoid contact of cleansing agents (like soap and alcohol)
with the filter because leakage may occur from the air
eliminating vent.
• Do not tape over filter(s) as this could block the air vent and
impede the infusion.
• Do not immerse pump in water. Take care to protect the
pump during any activities, which could cause pump and
filter to get wet, such as showering.
• In the event of any leakage from pump or administration
set, close tubing clamp. Replace pump if necessary.
• Do not discard the pump and contact Halyard Health for
product return instructions.
• Do not add unvented filter to end of the administration set
as this may impede or stop the flow rate.
• Do not remove the red tab until the tubing is completely
primed. Up to 5 ml bolus of air may be delivered if not
primed correctly.
•
Flow rates may vary due to:
Fill volume
• Filling pump less than the labeled volume results in
faster flow rate.
• Filling pump greater than the labeled results in slower
flow rate.
• Viscosity and/or drug concentration
• Pump position
- position pump at approximately the
same level as the catheter site:
• Positioning pump above this level increases flow rate.
• Positioning pump below this level decreases flow rate.
Temperature
• The ONDEMAND* device should be worn outside
clothing and kept at room temperature.
• To ensure flow rate accuracy, do not place heat or cold
therapy in close proximity to the flow controller.
• Temperature will affect solution viscosity, resulting in
faster or slower flow rate.
• ONDEMAND* device have been calibrated using Normal
Saline (NS) as the diluent and room temperature (22°C,
72°F) as the operating environment. Flow rate will
increase approximately 1.4% per 1°F/0.6°C increase in