4
Caution:
Do not fill less than the labeled fill volume or
exceed the maximum fill volume. (Table 1)
Table 1: Fill Volume
Labeled
Fill Vol
Maximum Fill Vol.
Retained
Vol.
270 ml
335 ml
≤ 9 ml
400 ml
550 ml
≤ 15 ml
Total Flow Rate Dosing
Total Flow Rate refers to bolus + basal, which is the infusion
rate per hour. (Table 2)
Table 2: Total Flow Rate
Bolus
Dose
Refill
Time
Total Flow Rate
5 ml
30 min
10 ml/hr + Basal Rate
5 ml
60 min
5 ml/hr + Basal Rate
Priming The Administration Set
Use Aseptic Technique
Caution:
Do not remove the red tab until the tubing is
completely primed. Up to a 5 ml bolus of air may be delivered
if not primed correctly.
1. Lay the ONDEMAND* device on flat surface with the red
tab label side up.
2. Open clamp and remove tubing cap to begin priming. Do
not discard tubing cap.
3. When all air has been removed from the entire tubing
and fluid is observed at end of Luer lock (approximately
4 minutes), the administration set is primed.
4. Replace tubing cap until ready for use.
5. Remove the red tab by pulling straight out (Figure 3-A). It
is important to remove red tab completely and ensure it
does not break (Figure 3-B). The ONDEMAND* bolus device
will begin to fill.
Instructions for Use
Use Aseptic Technique
Filling the ON-Q* Pump: (Figure 2)
Note:
Follow hospital protocols and applicable regulations for
filling pump.
1. Close clamp.
2. Uncap the fill port.
3. Attach filled syringe to fill port. Invert pump as shown.
4. Grasp syringe with both hands.
5. Push down on plunger continuously until volume is
dispensed. Do not handle pump while filling, as the
syringe tip may break. Repeat as necessary. Syringe
accuracy is ±4%.
Note:
Filling Extension Sets are provided with larger pumps
(see product insert).
Caution:
Do not underfill pump. Underfilling pump may
significantly increase the flow rate. Do not exceed maximum
fill volume. (Table 1)
6. Remove syringe from fill port.
7. Replace fill port cap. Label with the appropriate
pharmaceutical and patient information.
Note:
The ON-Q* Pump contains a Carry Case for
holding pump.
Figure 2