Basic Operations
January 2023
•
Rev. 0
• Standard Devices User Manual
11
Basic Operations
This section covers the most common functions of your
device. Settings for more specialized needs can be found
after this section. Some information found in this section may
be repeated in other sections for ease of section navigation.
Viewing Live Data
The display screen has a number of options for displaying
live data. There are various ways to modify the display screen
functions, brightness, or orientation. Refer to
display setup
The display screen displays live data for all parameters
measured by the device simultaneously. Sensors measure
data 1000 times per second and the LCD display updates
10 times per second. The measured data is displayed in the
units designated by the device (
).
Interacting with the Device
The images to the right show the buttons of the device display.
These images compare the differences between a mass flow
controller, a pressure controller, and a liquid meter.
Below are the default button functions and their locations
on a given device. Pressing a button to highlight a reading
changes what is centered in the display.
•
Highlight pressure:
all devices use
button 1.
•
Highlight volumetric flow:
mass flow
and liquid devices use
button 4
. Pressure
devices do not measure volumetric flow.
•
Highlight mass flow:
mass flow devices use
button 5
.
Pressure and liquid devices do not measure mass flow.
•
Highlight temperature:
mass flow and
liquid devices use
button 2
. Pressure
devices do not measure temperature.
•
Changing the setpoint:
controllers use
button 3
. Meters and gauges do not have a
setpoint function. Refer to
for more
information on changing the setpoint.
•
Taring the device:
meters and gauges use
button 3
(
TARE FLOW
for meters and
TARE PRESS
for gauges). Pressure controllers use
button 2
.
Mass flow controllers and liquid controllers do not
have a tare button on the default display. Refer to
for further information on taring.
•
Enter the menu or move to the next
screen:
all devices use
button 6.
•
Enable/Disable the backlight:
all devices use
button 7
.
Some devices don't use all 7 buttons. If an unused button is
pressed, nothing happens. Button functions can be modified
under
display setup
(
).
Status Messages
Status messages are shown to the right of the highlighted
central measurement:
ADC
Analog-digital converter error
EXH
Exhaust mode active
(Controllers only)
HLD
Valve hold active
(Controllers only)
LCK
Front display is locked
MOV
Mass flow over range of
mass flow device
OVR
Totalizer rolled over to zero
POV
Pressure over range of device
TMF
Totalizer missed out-of-range flow
TOV
Temperature over range of device
VOV
Volumetric flow over
range of device
Main display
of a
pressure
controller
measuring
gauge pressure.
Main display
of a
mass
flow controller
measuring
the mass flow of air.
Main display
of a
liquid
meter
measuring the
volumetric flow of water.