Device Setup
January 2023
•
Rev. 0
• Standard Devices User Manual
24
Creating New Mixes in COMPOSER™
MENU
→
SETUP
→
Active Gas
→
COMPOSER Mixes
→
Create Mix
Give the Mix a Short and Long Name
UP
/
DOWN
changes the character. Valid characters include A–Z,
0–9, punctuation (
. , -
), and space.
CANCEL
exits to the
mix
settings menu
.
SET
accepts the name.
✓
Note:
Using a space in the short name
can cause the serial data frame to be
read incorrectly by some programs.
Define the Mix.
•
Add Gas to Mix
enters the Gas Select™ category
listing. Once you find the correct gas, press
SET
.
Enter the composition percentage and press
SET
.
•
As gases are added, the total used percentage
updates on the
mix settings menu
.
•
Once gases have been added, COMPOSER™ can
change the gas percentage to fill the remaining
portion to 100% by selecting the component
gas, then selecting
Set % to Balance
.
•
The sum percentage of gases must total
100% to select
Save Mix
. Selecting
BACK
will permanently discard the mix.
•
Mixes that contain several gases push
the menu to a second page; use the
PAGE
button to see the remaining list.
Viewing, Deleting, and Creating Similar Mixes
MENU
→
SETUP
→
Active Gas
→
COMPOSER
Mixes
→
[Select mix]
→
INFO
The current configuration of any existing COMPOSER™ mix
can be viewed by selecting
INFO
instead of
SET
in the mix
list. It will show:
•
Options to delete the mix.
•
Create a similar mix.
•
Short and long names.
•
The gas number.
•
The composition, which may extend to a second page.
Pressing the
PAGE
button will move to the next page
Sensor Setup
MENU
→
SETUP
→
Sensor
Sensor setup controls how measurements are calculated
and communicated by the device. These are factors like
what engineering units are used and the standard or normal
reference points for flow.
Engineering Units
MENU
→
SETUP
→
Sensor
→
Engineering Units
Changing engineering units alters both the display and the
data frame. Choose the parameter whose unit you want to
change, and then choose an engineering unit, confirming the
change on the last screen.
Sensor setup menu
.
Engineering units menu
.
COMPOSER™ menu
without existing mixes.
Setting the percentage of
a constituent gas C2H2.
Results of adding of C2H2.
COMPOSER™ menu
with
the new custom mix.
Mix settings menu
.
Defining a mixture’s long name.