2
When a CO
2
/LN
2
backup system is activated, the installation place must be well ventilated.
Increased CO
2
in the air may be harmful and even fatal. If the ventilation is poor, other methods
should be considered in order to ensure safe working environments.
If there is a leakage of petroleum gas or other fl ammable gas, close the gas supply valve and
open doors and windows to ventilate the air. Do not plug or unplug your freezer unit in order to
avoid potential explosion or fi re.
Only professional technicians or Haier service personnel can install the unit. Failure to do so
may cause electricity or fi re.
The freezer must be securely installed on a fi rm fl oor. Tilted installation may result in the product
tipping over thereby causing injury and damage.
Please use a dedicated power supply matching the requirement specifi ed on the label to avoid
fi re and shock.
If the voltage being used is 10% higher than the rated voltage, a regulator with a capacity of
4000 W or higher must be installed.
If the power cord needs to be extended, the cross-section of the extended cable must be no less
than 2 mm² and no longer than 3 m for products of 208V~230V/50Hz or 208V~230V/60Hz and
no less than 3 mm² and no longer than 3 m for products of 115V~/60Hz to avoid fi re or electric
shock.
Your Haier ULT unit is equipped with a standard three-prong power plug(grounded)
complying with the standard three-prong socket (grounded) rated 16 A (208V~230V/50Hz or
208V~230V/60Hz) or rated 20A (115V~/60Hz). Removal of the ground prong is strictly prohibited
for safety reasons under any circumstances. The electrical power plug should be securely
plugged into the socket. A loose plug in the socket may cause fi re.
The power socket intended for your Haier ULT usage must be grounded to avoid electric shock.
If the socket does not meet this requirement, the condition must be corrected by a qualifi ed
technician before using the ULT unit.
Replacement of any spare parts should be performed by qualifi ed technicians.
Never install your ULT in an unprotected area. If the unit is rained on, there is a danger of
electric shock.
Your Haier ULT must not be installed in a damp area or an area subjected to water spray.
Otherwise this may reduce the degree of insulation and thereby cause electrical leakage or
electrical shock.
Never directly pour water into the unit. The water may cause electric shock or short circuit.
Do not place any water container or heavy object on top of the unit. A falling object may injure
an operator. If the water spills into the unit, it may damage the insulation thereby causing electric
shock.
Never use gas lines, water mains, telephone lines or lightening rods as the grounding device for
your Haier ULT unit. This type of improper grounding may cause electric shock or other danger.
Do not touch any electrical components, switches or power cord with wet hands. Such action
may lead to electric shock.
47
Technical Data
Model
Net
Volume
(L)
Rated
Voltage
(VAC)
Rated
Frequency
(Hz)
Rated
Power
(W)
Weight
(kg)
Dimensions
(W×D×H)
(mm)
DW-86L579BP
579
100-230~
50/60
1100
320
895×998×1980
DW-86L729BP
729
100-230~
50/60
1100
350
1046×998×1980
DW-86L579BPT
579
100-230~
50/60
1100
320
895×998×1980
DW-86L729BPT
729
100-230~
50/60
1100
350
1046×998×1980
DW-86L829BP
829
208-230~
50/60
1100
380
1145×998×1980
DW-86L959BP
959
208-230~
50/60
1300
450
1296×998×1980
DW-86L829BPT
829
208-230~
50/60
1100
380
1145×998×1980
DW-86L959BPT
959
208-230~
50/60
1300
450
1296×998×1980
DW-86L959W
959
208-230~
50/60
1300
450
1296×998×1980
DW-86L828W
828
208-230~
50/60
1100
380
1145×998×1980
Packing List
Name
User
Manual
Instruction to
install sapcers
Plastic
bag
Ice
Scraper
Key
Spacer
DW-86L579BP
1
1
1
1
4
2
DW-86L729BP
1
1
1
1
4
2
DW-86L579BPT
1
1
1
1
4
2
DW-86L729BPT
1
1
1
1
4
2
DW-86L829BP
1
1
1
1
4
2
DW-86L959BP
1
1
1
1
4
2
DW-86L829BPT
1
1
1
1
4
2
DW-86L959BPT
1
1
1
1
4
2
DW-86L959W
1
1
1
1
4
2
DW-86L828W
1
1
1
1
4
2