Describing the MCS+ Cabinet Components
2-5
P/N 85213-30, Manual revision: B
Understanding the function of the pumps
The pumps function in a general manner during the different modes of MCS+
operation as follows:
When loading the disposable tubing:
!
The AC pump and the Blood pump turn simultaneously to thread the dis-
posable tubing onto the pump rotors.
!
The Transfer pump will be active during certain MCS+ protocols.
During PRIME
!
The AC pump and the Blood pump turn simultaneously to provide the inlet
side of the donor-line tubing with AC solution.
!
The Transfer pump will be active during certain MCS+ protocols.
During DRAW
!
The AC pump will deliver AC solution from the AC solution bag to the nee-
dle connector.
!
The Blood pump will pull anticoagulated whole blood past the blood filter
of the disposable set and into the centrifuge bowl.
!
The Transfer pump will be active during certain protocols. The Transfer
pump speed will vary in relation to the Blood pump speed and the donor
flow rate.
During RETURN
!
The Blood pump pulls the remaining blood components from the centri-
fuge bowl and re-infuses the contents to the donor.
!
The Transfer pump will be used to mix solutions with the uncollected
blood components, prior to re-infusion to the donor.
!
The AC pump will be inactive.
Note: The Blood pump and the AC pump will rotate at different speeds during
DRAW, depending on the AC/Blood pump ratio parameter setting. Information
related to the function of the pumps during sub states, such as SURGE and
DWELL, will be provided in each relevant MCS+ protocol manual.
Table 2-1, MCS+ pump function according to MCS+ operating mode
MCS+ mode
AC pump
Transfer pump
Blood pump
PRIME
Yes
Protocol-specific
Yes
DRAW
Yes
Protocol-specific
Yes
RETURN
No
Protocol-specific
Yes
Summary of Contents for MCS+
Page 2: ...2002 2006 Haemonetics Corporation All rights reserved...
Page 10: ...x Preface P N 85213 30 Manual revision B The Haemonetics MCS LN09000 220E ED...
Page 20: ......
Page 30: ......
Page 81: ...Maintaining the MCS Equipment 5 7 P N 85213 30 Manual revision B...
Page 85: ......
Page 86: ...5 12 Maintaining the MCS Equipment P N 85213 30 Manual revision B...
Page 100: ......