xviii
Preface
P/N 85213-30, Manual revision: B
Specifications of
the MCS+ device
The approximate weight and dimensions for the MCS+ device are as follows:
The following environmental conditions should be respected pertaining to oper-
ation and storage of the MCS+ device:
The electrical specifications for operating the MCS+ device are as follows:
Note: Haemonetics will regulate the proper voltage setting upon installation. The
power source used must be properly grounded.
Warning: Equipment not suitable for use in the presence of a flammable anes-
thetic mixture with air or with oxygen or nitrous oxide.
The shipping/ storage temp are -20C to 50C at 8% to 80% humidity. The
artwork for the carton is only in the old film style.
Characteristics
Values
Cabinet
cover open
Cabinet
cover closed
Height
68.5 cm
44 cm
Width
56.5 cm
Depth
56.5 cm
36.5 cm
Depth with communication box
56.5 cm
38.5 cm
Weight
27.5 kg
Weight with communication box
28.5 kg
Conditions
Values
Ambient operating temperature
+18° C to +27° C
Tested storage temperature
-20° C to +50° C
Storage humidity level
8% to 80%, non-condensing
Characteristics
Values
(relative to input voltage)
Input voltage
230 VAC ± 10%
110 VAC ± 10%
Operating current
~1.9 A
~ 2.6 A
Fuse rating
F2.5 A @ 250 V
F5.0 A @ 250 V
Operating frequency range
50 - 60 Hz
50 - 60 Hz
Summary of Contents for MCS+
Page 2: ...2002 2006 Haemonetics Corporation All rights reserved...
Page 10: ...x Preface P N 85213 30 Manual revision B The Haemonetics MCS LN09000 220E ED...
Page 20: ......
Page 30: ......
Page 81: ...Maintaining the MCS Equipment 5 7 P N 85213 30 Manual revision B...
Page 85: ......
Page 86: ...5 12 Maintaining the MCS Equipment P N 85213 30 Manual revision B...
Page 100: ......