5-2
Maintaining the MCS+ Equipment
P/N 85213-30, Manual revision: B
C
LEANING
PROCEDURES
The MCS+ device has been designed to require minimal maintenance for the
operator. To maintain the precision function of the MCS+ device, the operator
needs to primarily perform routine cleaning procedures of certain key compo-
nents. A record of routine cleaning schedule can be kept along with any routine
or preventive service maintenance performed by a Haemonetics representative.
The frequency of cleaning each individual MCS+ device will depend on the
number of procedures performed. Special cleaning needs may arise and should
be dealt with promptly. Haemonetics recommends the following routine
cleaning schedule for each MCS+ device, based on an average of three collec-
tion procedures per day, or approximately sixty per month.
!
Daily: Clean the exterior surfaces as well as the pressure monitors.
!
Weekly: Clean the air detectors, the optical sensors (line sensor and optical
bowl sensor), the fluid detector, and the inside of the centrifuge well.
!
Monthly: Clean the pump rotors and the pump wells.
!
Quarterly: Clean the filter screens.
Warning: To eliminate the potential danger of electrical shock, the operator
must clean the MCS+ device only when it is disconnected from an external pow-
er source.
The following list describes the basic material required for routine cleaning.
!
Disinfectant cleaning solution, specific for blood-born pathogens.
!
Warm water.
!
70% Isopropyl alcohol.
!
Lint-free gauze or cloth (for cleaning and drying).
!
Cotton swabs.
!
Protective gloves.
!
Hexagonal-head wrench #10 (provided with the device).
!
Silicon lubricant (for the “L” gasket of the centrifuge vacuum chuck).
!
Phillips-head screwdriver.
Cabinet and
control panel
The exterior cabinet, valves, keypad and display screen should be wiped daily,
(as well as following any spill), using the cleaning solution.
Summary of Contents for MCS+
Page 2: ...2002 2006 Haemonetics Corporation All rights reserved...
Page 10: ...x Preface P N 85213 30 Manual revision B The Haemonetics MCS LN09000 220E ED...
Page 20: ......
Page 30: ......
Page 81: ...Maintaining the MCS Equipment 5 7 P N 85213 30 Manual revision B...
Page 85: ......
Page 86: ...5 12 Maintaining the MCS Equipment P N 85213 30 Manual revision B...
Page 100: ......