9.3.4
Transition
The Figure 9-7 shows the “transition” menu. It is possible to select independently “Duration”, “Interval”, “Phase” and
“Alternate polarity”. To be able to select “phase” the Synchrony has to be either set as Synchronization or external
Synchronization, otherwise it is not possible to access it. Eventually, every parameter can be changed by pushing the
certain value.
Figure 9-7: Voltage Dips Transition
When finalizing the parameter, it must be confirmed with “OK” and either the pre-compliance or standard mode gets
open depending what selection was previously done. In addition the incremental intervals get displayed at each
particular function name, like for example “interval”.
9.3.5
Trigger
The trigger can be accessed either from pre-compliance mode or standard mode when generator is not in use. The
user can select after pressing the “trigger” button between: “auto”, “manual” and “external”. In “Auto”: signal gets
applied automatically in accordance to parameters. In “Manual”: pushing the start button after “manual” selection, then
pressing the “trigger” button. In “External”: Signal will be controlled from external source, connection through PIN 11 in
Table 5-3.