AXOS
5
Compact immunity test System
19
7
Magnetic field
7.1
General Information
The test setup is described in accordance to the norm IEC 61000-4-9. A SCC waveform will be created. For the
magnetic field the coil gets connected directly via banana plug to the surge output HI and COM as written in position 6
and Position 7 in Figure 5-1. Significant is that only these outputs can be used for the Surge impulse.
Do not touch the coil when the test is proceeding. Do fulfil safety requirements as described in capture 1.1.
In Figure 7-1 is the pre-compliance mode presented. The parameter can be entered by pressing the numbers. For
instance to change the magnet field the parameter “+120 A/m” must be selected and then e new parameter entered.
Figure 7-1: Magnetic field Pre-compliance mode
In addition, the purple bow describes the function. On the top of the window it shows which “Coil Antenna” is selected.
In the standard mode (Figure 7-2) the values can be changed by pressing on the particular parameter as well.
Additionally, the “transition” (capture 8.2.3) can be adjusted.
Figure 7-2: Magnetic field Standard mode
When “Line” button pushed, “EUT Supply Input” connected with L, N, PE front view and in the right corner the plug
flashes red. After pushing the “Start” button the surge impulse gets applied.