CAUTION Use the Smartdop
®
30EX under normal operating conditions, according to the instructions
provided in this operation manual and under the operating conditions stated in the product
specifications.
CAUTION Federal law restricts this device to sale by or on the order of a licensed healthcare
practitioner.
CAUTION The Smartdop
®
30EX has not been evaluated for safety and compatibility in the MR
environment. The Smartdop
®
30EX has not been tested for heating or migration in the MR
environment.
CAUTION Perform ultrasound testing/procedures prudently using the principle of ALARA (As Low As
Reasonably Achievable).
CAUTION Use only the designated adaptor that came with your Smartdop
®
30EX or contact Koven
Technology for parts & accessories.
CAUTION To avoid possible damage to the printer, never pull paper in the opposite direction as the
paper feed. To properly remove paper, cut the paper at the roll side and pull the remaining
paper out through the top of the unit.
CAUTION Use ultrasound gel. Other materials such as baby oil or cream may produce incorrect
sounds or damage the probe.
WARNING The ultrasound gel included with purchase is non-sterile and should only be used
noninvasively.
WARNING Discontinue use of ultrasound gel if an allergic reaction occurs.
CAUTION The probe transducer tip is thin and delicate. Please handle with care and use the probe cap
when not in use.
WARNING If any abnormality is found on the unit or patient, discontinue use in a manner safe to the
patient.
WARNING In case of emergency or unexpected situations, stop cuff inflation by pressing the INF/DUMP
button on the front panel, disconnect the tubing from the unit, and turn the unit off.
CAUTION Make sure all cables are property connected and there are no obstructions to any external
connections to the patient, before use.
CAUTION Do not drop or place in an area where the unit may fall.
CAUTION Do not use simultaneously with an electrocautery defibrillator, other ultrasonic device or a
mobile phone.
CAUTION Do not use in an area where a strong electromagnetic field is present. Doing so may cause
incorrect measurements.
CAUTION Do not place near water or other liquid, where atmospheric pressure, temperature, humidity,
ventilation, sunlight, dust, salt sulfur, etc. may cause damage or where chemicals are stored
or gas may be generated.
CAUTION Protect the unit from vibration and shock during transport.
WARNINGS & CAUTIONS