Section 1 Specifications
Specifications are subject to change without notice. The specifications that follow are for the junction
box (
). Refer to the instrument user manuals for the
analyzer, controller, turbidimeter and probe sensor specifications.
Table 1 Junction box
Specification
Details
Power requirements
100 to 115/230 VAC, 50/60 Hz
Power consumption
90 VA maximum for CL17 analyzer
24 W maximum for all other electronics
Protection class
I
Installation category
II
Pollution degree
2
Altitude
2000 m (6562 ft) maximum
Fuse
5 x 20 mm, T 1.0 A, 250 V (2x)
Storage temperature –20 to 60 °C (–4 to 140 °F)
Conduit hub
½ NPT
Communications
RS485 Modbus
Humidity
Maximum relative humidity 80% for temperatures to 31 °C (88 °F), decreasing linearly to
50% relative humidity at 40 °C (104 °F)
Table 2 WDMP sc panel
Specification
Details
Dimensions (W x H x D)
81 x 122 x 30.8 cm (32 x 48 x 12.1 in.)
Weight
54.4 kg (120 lb)
Operating temperature
5 to 40 °C (41 to 104 °F)
Sample flow rate
0.4 to 0.6 L/min
Sample pressure
138 to 862 kPa (20 to 125 psi)
Sample temperature
5 to 40 °C (41 to 104 °F)
Sample inlet connection
½-in. OD tube
Sample waste drain connection ¾-in. ID hose barb
Sample waste drain pressure
Ambient, free-flowing (atmospheric)
Pressure sensor
Stainless steel (17-4 PH), Input: 7–35 VDC; Output: 4–20 mA, 0–1034 kPa
(0–150 psi)
English
3
Summary of Contents for WDMP sc
Page 2: ...Table of Contents English 3 Español 24 Français 48 2 ...
Page 32: ...Figura 4 Dimensiones del producto 1 Espacio libre 32 Español ...
Page 36: ...2 Abra la caja de conexiones Consulte la Figura 7 36 Español ...
Page 56: ...Figure 4 Dimensions du produit 1 Zone tampon 56 Français ...
Page 72: ......
Page 73: ......