Figure 9 Install the probes
1
Probe adapter
5
Sensor port for optional ORP probe
2
Protective cap
6
Standard conductivity probe
3
pH probe
7
Connections on the junction box
4
Sensor port plug
3.3.3 Prepare the CL17 chlorine analyzer
The CL17 chlorine analyzer comes with a startup kit and reagents. Refer to the CL17 user manual to
complete the tasks that follow.
• Install the stir bar in the colorimeter.
• Prepare the DPD indicator reagent.
• Install the reagents.
• Install the pinch plate on the pump/valve module.
3.3.4 Connect the sample and drain lines
The sample line connects to a ½-inch OD quick-connect tube fitting. The supplied drain tubing
connects to a ¾-inch barb fitting.
1.
Install a pressure regulator in the sample line if the sample pressure is more than 862 kPa
(125 psi).
2.
Push to install the sample line in the quick-connect fitting below the sample flow valve. Refer to
3.
Use the supplied drain tubing to connect the panel drain to the external drains. Refer to
. Make sure that the drain tubing has a constant downward slope.