49
SECTION 5
DISSOLVED OXYGEN OPERATION
MODE
5.1
Dissolved Oxygen Setup Menu Options
Table 10 describes the options available in the Dissolved Oxygen
Setup Menu.
Salinity Factor — This feature adjusts the displayed dissolved
oxygen concentration in mg/L based on the sample’s salinity.
When the
Sal
icon is displayed during the reading mode, a salinity
correction calculation is applied to the dissolved oxygen
concentration in mg/L. The dissolved oxygen concentration in
% saturation is the ratio of the displayed concentration in mg/L to
the equilibrium dissolved oxygen concentration for the sample’s
temperature and salinity plus ambient barometric pressure.
5.2
How to Change the Dissolved Oxygen Menu Options
To access the Dissolved Oxygen Setup menu:
1. Turn on the meter and press the
DO
key. The arrow icons that
appear indicate that additional options are available within
the menu.
2. Press the
SETUP
key.
3. Use the
UP
ARROW
and
DOWN ARROW
keys to scroll
between the desired options.
Table 10 Setup Options
Setup
Number
Setup Description
Option Description
Default Setting
1
Time
00:00 to 23:59
00:00
2
Date
01/01 to 12/31
01/01
3
Year
2000–2099
2000
4
Temperature units
°C, °F
°C
5
Display lock
On:
Lock
icon
Off:
Lock
icon and
(Off)
icon
Off
6
Resolution
0.0, 0.00
0.00
7
Salinity Factor
0–40
0
8
Auto-print interval
Off, 10 sec., 30 sec., 1 min., 5 min., 20 min.,
1 hr., 2 hrs., and 6 hrs.
Off