28
SECTION 3,
continued
Accessing the Calibration mode or the Setup mode halts
automatic data transfer. Also, if the meter has been set to the Lock
mode using Setup 1, the meter will not send data. When the meter
is in Lock mode and the
READ
key is pressed, automatic data
transfer will occur at selected time intervals only until the meter
stabilizes and the value in the display is locked.
To keep transferring pH data, leave the meter in pH Reading
mode on a docking station.
Each time data transfer occurs, the Print icon will momentarily
appear at the top of the display.
When Auto Print is on, the instrument will send data at the
specified interval, as long as the instrument is in the pH Reading
mode.
3.2
How to Change the pH Menu Options
To access the pH Setup menu:
1. Turn on the meter and press the
p
H
key.
2. Press the
SETUP
key.
3. Use the
UP
ARROW
and
DOWN ARROW
keys to scroll
between the desired options.
Table 4 shows how to change each specific setup option.