
Troubleshooting
Problem
Possible cause
Solution
The instrument does not
operate. The LED status
indicators are off.
The power switch is off.
Set the power switch to on.
The power cord is disconnected
from the power supply or wall
outlet.
Connect the power cord.
The power cord is damaged.
Replace the power cord.
The power supply is faulty.
Contact technical support.
The electrical transients on the
power line are too large, which
may disrupt the USB
communication between the
analyzer and the sampler.
Install any commercially available USB
opto-isolator between the analyzer and the
auto sampler. To repair USB
communications, set the power for both
instruments to off, then to on.
The instrument does not
operate. The LED status
indicators are on.
The home position of the arm
has damage.
Contact technical support.
The sample probe is not moving
freely.
Remove the probe to clean and remove
debris. Refer to
on page 15.
Replace the sample probe. Refer to
on page 15.
The mechanical Z-drive has
damage.
Contact technical support.
The mechanical Z-drive is not
installed correctly.
Validate the installation. Refer to
on page 9.
The arm does not operate
correctly.
The instrument was physically
damaged.
Set the instrument power to off, then to on.
If the problem continues, contact technical
support.
The sample probe does not pull
up the sample.
The air flow into vial does not
correctly flow. The sample
tubing is loose.
Examine the sample probe. Make sure that
the needle is inside the metal sleeve.
Tighten the sample tubing connection.
The sample probe does not go
through the septum cap of the
vial.
The stripper plate is not aligned
correctly.
Adjust the height between the vials in the
sample tray and the platform of the stripper
plate. Refer to
The probe does not easily go
into the mechanical Z-drive.
The sample bottle is lifted up
after the probe retracts.
Replacement parts and accessories
W A R N I N G
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
16
English
Summary of Contents for QbD1200 AutoSampler
Page 2: ...English 3 Español 18 Français 33 Português 48 日本語 63 한글 78 中文 93 中文 107 2 ...
Page 71: ...配管 TOC 分析装置への配管 次の図示された手順を参照して TOC 分析装置から試料チューブをオートサンプラーの機械式 Z ド ライブに接続します 日本語 71 ...
Page 86: ...배관 분석기 배관 연결 분석기의 샘플 튜브를 자동 샘플 채취기의 기계적 Z 드라이브에 연결하려면 아래 그림의 순서를 참조 하십시오 86 한글 ...
Page 100: ...安装针套和取样针 警 告 刺伤危险 暴露的针头可能导致刺伤 小心安装或移动瓶身 请参阅以下图示步骤安装针套和取样针 100 中文 ...
Page 101: ...装设管道 连接至分析仪 请参阅以下图示步骤 将样品管从分析仪连接至自动采样器上的机械 Z 驱动 中文 101 ...
Page 114: ...安裝套筒和樣品探針 警 告 有刺傷的危險 露出的針頭可能會導致穿刺傷 在安裝或移除瓶子時 請謹慎小心 請參閱下列圖示步驟所示來安裝針頭套筒和樣品探針 114 中文 ...
Page 115: ...配管 連接管線至分析儀 請參閱下列圖示步驟所示 將分析儀的取樣管連接至自動取樣器上的機械 Z drive 驅動裝置 中文 115 ...
Page 121: ......