8.
Review the calibration result:
• Passed—the sensor is calibrated and ready to measure samples.
The slope and/or offset values are shown.
• Failed—the calibration slope or offset is outside of accepted limits.
Repeat the calibration with fresh reference solutions. Refer to
on page 15 and
on page 16 for
more information.
9.
If the calibration passed, push
ENTER
to continue.
10.
If the option for operator ID is set to Yes in the Calibration Options
menu, enter an operator ID. Refer to
on page 14.
11.
On the New Sensor screen, select whether the sensor is new:
Option Description
Yes
The sensor was not calibrated previously with this controller. The
days of operation and previous calibration curves for the sensor are
reset.
No
The sensor was calibrated previously with this controller.
12.
Return the sensor to the process and push
ENTER
.
The output signal returns to the active state and the measured
sample value is shown on the measure screen.
Note: If the output mode is set to hold or transfer, select the delay time when
the outputs return to the active state.
Temperature calibration
The instrument is calibrated at the factory for accurate temperature
measurement. The temperature can be calibrated to increase accuracy.
1.
Put the sensor in a container of water.
2.
Measure the temperature of the water with an accurate thermometer
or independent instrument.
3.
Push the
MENU
key and select Sensor Setup, [Select Sensor],
Calibrate.
4.
Select 1 PT Temp Cal and push
ENTER
.
5.
Wait for the value to stabilize and push
ENTER
.
6.
Enter the exact value and push
ENTER
.
7.
Return the sensor to the process and push
ENTER
.
Exit calibration procedure
If the
BACK
key is pushed during a calibration, the user can exit the
calibration.
1.
Push the
BACK
key during a calibration. Three options are shown:
Option
Description
QUIT CAL
Stop the calibration. A new calibration must start from the
beginning.
BACK TO CAL
Return to the calibration.
LEAVE CAL
Exit the calibration temporarily. Access to other menus is
allowed. A calibration for a second sensor (if present) can
be started. To return to the calibration, push the
MENU
key
and select Sensor Setup, [Select Sensor].
2.
Use the arrow keys to select one of the options and push
ENTER
.
Change calibration options
The user can set a reminder or include an operator ID with calibration
data from the CAL OPTIONS menu.
1.
Push the
MENU
key and select Sensor Setup, [Select Sensor],
Calibrate, Cal Options.
2.
Use the arrow keys to select an option and push
ENTER
.
Option
Description
CAL REMINDER
Sets a reminder for the next calibration in days, months or
years—Off (default), 1 day, 7, 30, 60, or 90 days, 6 or
9 months, 1 or 2 years
OP ID on CAL
Includes an operator ID with calibration data—Yes or No
(default). The ID is entered during the calibration.
14
English
Summary of Contents for Polymetron
Page 2: ...English 3 Français 20 Español 38 Português 58 中文 76 日本語 92 한글 109 ไทย 125 2 ...
Page 8: ...1 2 3 4 8 English ...
Page 9: ...5 6 7 8 English 9 ...
Page 25: ...1 2 3 4 Français 25 ...
Page 26: ...5 6 7 8 26 Français ...
Page 42: ...Figura 3 Ejemplos de montaje 1 3 4 pulg NPT 2 1 4 pulg NPT 42 Español ...
Page 44: ...1 2 3 4 44 Español ...
Page 45: ...5 6 7 8 Español 45 ...
Page 63: ...1 2 3 4 Português 63 ...
Page 64: ...5 6 7 8 64 Português ...
Page 79: ...图 3 安装示例 1 流通室 T 型安装 聚氯乙烯 英寸 NPT 2 流通室 T 型安装 不锈钢 1 4 英寸 NPT 中文 79 ...
Page 81: ...1 2 3 4 中文 81 ...
Page 82: ...5 6 7 8 82 中文 ...
Page 95: ...図 3 取り付け例 1 フロースルー T 字取り付け PVC 19 mmNPT 2 フロースルー T 字取り付け ステンレス鋼 6 mmNPT 日本語 95 ...
Page 97: ...1 2 3 4 日本語 97 ...
Page 98: ...5 6 7 8 98 日本語 ...
Page 112: ...그림 3 장착 예 1 흐름 통과 T 장착부 PVC 3 4 인치 NPT 2 흐름 통과 T 장착부 스테인리스 스틸 1 4 인치 NPT 112 한글 ...
Page 114: ...1 2 3 4 114 한글 ...
Page 115: ...5 6 7 8 한글 115 ...
Page 130: ...1 2 3 4 130 ไทย ...
Page 131: ...5 6 7 8 ไทย 131 ...
Page 142: ......
Page 143: ......