11.
On the New Sensor screen, select whether the sensor is new:
Option Description
Yes
The sensor was not calibrated previously with this controller. The
days of operation and previous calibration curves for the sensor are
reset.
No
The sensor was calibrated previously with this controller.
12.
Proceed to the calibration with a reference solution or process
sample.
Calibration with a reference solution
Calibration adjusts the sensor reading to match the value of a reference
solution. Use a reference solution that is at the same value or higher
than the expected measurement readings.
Note: If the sensor is being calibrated for the first time, be sure to complete the
zero calibration first.
1.
Thoroughly rinse the clean sensor in deionized water.
2.
Put the sensor in the reference solution. Support the sensor so that it
does not touch the container. Make sure that the sensing area is fully
immersed in the solution (
). Stir the sensor to remove
bubbles.
Figure 4 Sensor in reference solution
3.
Wait for the sensor and solution temperature to equalize. This can
take 30 minutes or more if the temperature difference between the
process and reference solution is significant.
4.
Push the
MENU
key and select Sensor Setup, [Select Sensor],
Calibrate.
5.
Select the calibration for the specified parameter and push
ENTER
:
• Conductivity—Cond Cal
• TDS—TDS Cal
• Salinity—Cond Cal
• Concentration—Conc Cal or Cond Cal
6.
If the passcode is enabled in the security menu for the controller,
enter the passcode.
7.
Select the option for the output signal during calibration:
Option
Description
Active
The instrument sends the current measured output value during
the calibration procedure.
12
English
Summary of Contents for Polymetron
Page 2: ...English 3 Français 20 Español 38 Português 58 中文 76 日本語 92 한글 109 ไทย 125 2 ...
Page 8: ...1 2 3 4 8 English ...
Page 9: ...5 6 7 8 English 9 ...
Page 25: ...1 2 3 4 Français 25 ...
Page 26: ...5 6 7 8 26 Français ...
Page 42: ...Figura 3 Ejemplos de montaje 1 3 4 pulg NPT 2 1 4 pulg NPT 42 Español ...
Page 44: ...1 2 3 4 44 Español ...
Page 45: ...5 6 7 8 Español 45 ...
Page 63: ...1 2 3 4 Português 63 ...
Page 64: ...5 6 7 8 64 Português ...
Page 79: ...图 3 安装示例 1 流通室 T 型安装 聚氯乙烯 英寸 NPT 2 流通室 T 型安装 不锈钢 1 4 英寸 NPT 中文 79 ...
Page 81: ...1 2 3 4 中文 81 ...
Page 82: ...5 6 7 8 82 中文 ...
Page 95: ...図 3 取り付け例 1 フロースルー T 字取り付け PVC 19 mmNPT 2 フロースルー T 字取り付け ステンレス鋼 6 mmNPT 日本語 95 ...
Page 97: ...1 2 3 4 日本語 97 ...
Page 98: ...5 6 7 8 98 日本語 ...
Page 112: ...그림 3 장착 예 1 흐름 통과 T 장착부 PVC 3 4 인치 NPT 2 흐름 통과 T 장착부 스테인리스 스틸 1 4 인치 NPT 112 한글 ...
Page 114: ...1 2 3 4 114 한글 ...
Page 115: ...5 6 7 8 한글 115 ...
Page 130: ...1 2 3 4 130 ไทย ...
Page 131: ...5 6 7 8 ไทย 131 ...
Page 142: ......
Page 143: ......