Option
Description
Hold
The sensor output value is held at the current measured value
during the calibration procedure.
Transfer
A preset output value is sent during calibration. Refer to the
controller user manual to change the preset value.
8.
With the sensor in the reference solution, push
ENTER
.
9.
Enter the reference temperature of the reference solution and push
ENTER
.
10.
Enter the slope of the reference solution and push
ENTER
.
11.
Wait for the value to stabilize and push
ENTER
.
Note: The screen may advance to the next step automatically.
12.
Use the arrow keys to enter the value of the reference solution and
push
ENTER
.
13.
Review the calibration result:
• Passed—the sensor is calibrated and ready to measure samples.
The slope and/or offset values are shown.
• Failed—the calibration slope or offset is outside of accepted limits.
Repeat the calibration with fresh reference solutions. Refer to
on page 15 and
on page 16 for
more information.
14.
If the calibration passed, push
ENTER
to continue.
15.
If the option for operator ID is set to Yes in the Calibration Options
menu, enter an operator ID. Refer to
on page 14.
16.
On the New Sensor screen, select whether the sensor is new:
Option Description
Yes
The sensor was not calibrated previously with this controller. The
days of operation and previous calibration curves for the sensor are
reset.
No
The sensor was calibrated previously with this controller.
17.
Return the sensor to the process and push
ENTER
.
The output signal returns to the active state and the measured
sample value is shown on the measure screen.
Note: If the output mode is set to hold or transfer, select the delay time when
the outputs return to the active state.
Calibration with the process sample
The sensor can remain in the process sample, or a portion of the
process sample can be removed for calibration.
1.
Push the
MENU
key and select Sensor Setup, [Select Sensor],
Calibrate.
2.
Select the calibration for the specified parameter and push
ENTER
:
• Conductivity—Cond Cal
• TDS—TDS Cal
• Salinity—Cond Cal
• Concentration—Conc Cal or Cond Cal
3.
If the passcode is enabled in the security menu for the controller,
enter the passcode.
4.
Select the option for the output signal during calibration:
Option
Description
Active
The instrument sends the current measured output value during
the calibration procedure.
Hold
The sensor output value is held at the current measured value
during the calibration procedure.
Transfer
A preset output value is sent during calibration. Refer to the
controller user manual to change the preset value.
5.
With the sensor in the process sample, push
ENTER
.
The measured value is shown.
6.
Wait for the value to stabilize and push
ENTER
.
7.
Use the arrow keys to enter the value of the process sample and
push
ENTER
.
English
13
Summary of Contents for Polymetron
Page 2: ...English 3 Français 20 Español 38 Português 58 中文 76 日本語 92 한글 109 ไทย 125 2 ...
Page 8: ...1 2 3 4 8 English ...
Page 9: ...5 6 7 8 English 9 ...
Page 25: ...1 2 3 4 Français 25 ...
Page 26: ...5 6 7 8 26 Français ...
Page 42: ...Figura 3 Ejemplos de montaje 1 3 4 pulg NPT 2 1 4 pulg NPT 42 Español ...
Page 44: ...1 2 3 4 44 Español ...
Page 45: ...5 6 7 8 Español 45 ...
Page 63: ...1 2 3 4 Português 63 ...
Page 64: ...5 6 7 8 64 Português ...
Page 79: ...图 3 安装示例 1 流通室 T 型安装 聚氯乙烯 英寸 NPT 2 流通室 T 型安装 不锈钢 1 4 英寸 NPT 中文 79 ...
Page 81: ...1 2 3 4 中文 81 ...
Page 82: ...5 6 7 8 82 中文 ...
Page 95: ...図 3 取り付け例 1 フロースルー T 字取り付け PVC 19 mmNPT 2 フロースルー T 字取り付け ステンレス鋼 6 mmNPT 日本語 95 ...
Page 97: ...1 2 3 4 日本語 97 ...
Page 98: ...5 6 7 8 98 日本語 ...
Page 112: ...그림 3 장착 예 1 흐름 통과 T 장착부 PVC 3 4 인치 NPT 2 흐름 통과 T 장착부 스테인리스 스틸 1 4 인치 NPT 112 한글 ...
Page 114: ...1 2 3 4 114 한글 ...
Page 115: ...5 6 7 8 한글 115 ...
Page 130: ...1 2 3 4 130 ไทย ...
Page 131: ...5 6 7 8 ไทย 131 ...
Page 142: ......
Page 143: ......