Troubleshooting
Intermittent data
During calibration, data is not sent to the datalog. Thus, the datalog can
have areas where the data is intermittent.
Test the conductivity sensor
If a calibration fails, first complete the maintenance procedures in
on page 15.
1.
Disconnect the sensor wires.
2.
Use an ohmmeter to test the resistance between the sensor wires as
shown in
.
Note: Be sure that the ohmmeter is set to its highest range for all infinite (open
circuit) resistance readings.
Table 4 Conductivity resistance measurements
Measurement points
Resistance
Between blue and white wires
1089–1106 ohms at 23–27 °C
Between red wire and sensor body
Less than 5 ohms
Between black wire and inner electrode
Less than 5 ohms
Between black and red wires
Infinite (open circuit)
Between black and white wires
Infinite (open circuit)
Between red and white wires
Infinite (open circuit)
Between red and inner shield wires
Infinite (open circuit)
Between black and inner shield wires
Infinite (open circuit)
Between white and inner shield wires
Infinite (open circuit)
Between outer and inner shield wires
Infinite (open circuit)
If one or more of the measurements is incorrect, call technical support.
Supply technical support with the serial number of the sensor and the
resistance values measured.
Sensor diagnostic and test menu
The sensor diagnostic and test menu shows current and historical
information about the instrument. Refer to
diagnostic and test menu, push the
MENU
key and select Sensor Setup,
[Select Sensor], DIAG/TEST.
Table 5 Sensor DIAG/TEST menu
Option
Description
SENSOR INFORMATION Shows the name and serial number that was entered
by the user.
CARD INFORMATION
Shows the version and the serial number for the
sensor module.
CAL DAYS
Shows the number of days since the last calibration.
CAL HISTORY
Shows a list of the calibrations and the details for
each calibration.
RESET CAL HISTORY
Service use only. Resets the calibration history for
the sensor. All previous calibration data is lost.
SENSOR SIGNALS
Shows the current sensor signal and span in µS/cm.
SENSOR DAYS
Shows the number of days that the sensor has been
in operation.
RESET SENSOR DAYS
Resets the Sensor Days counter.
Error list
When an error occurs, the reading on the measurement screen flashes
and all outputs are held when specified in the controller menu. To show
16
English
Summary of Contents for Polymetron
Page 2: ...English 3 Français 20 Español 38 Português 58 中文 76 日本語 92 한글 109 ไทย 125 2 ...
Page 8: ...1 2 3 4 8 English ...
Page 9: ...5 6 7 8 English 9 ...
Page 25: ...1 2 3 4 Français 25 ...
Page 26: ...5 6 7 8 26 Français ...
Page 42: ...Figura 3 Ejemplos de montaje 1 3 4 pulg NPT 2 1 4 pulg NPT 42 Español ...
Page 44: ...1 2 3 4 44 Español ...
Page 45: ...5 6 7 8 Español 45 ...
Page 63: ...1 2 3 4 Português 63 ...
Page 64: ...5 6 7 8 64 Português ...
Page 79: ...图 3 安装示例 1 流通室 T 型安装 聚氯乙烯 英寸 NPT 2 流通室 T 型安装 不锈钢 1 4 英寸 NPT 中文 79 ...
Page 81: ...1 2 3 4 中文 81 ...
Page 82: ...5 6 7 8 82 中文 ...
Page 95: ...図 3 取り付け例 1 フロースルー T 字取り付け PVC 19 mmNPT 2 フロースルー T 字取り付け ステンレス鋼 6 mmNPT 日本語 95 ...
Page 97: ...1 2 3 4 日本語 97 ...
Page 98: ...5 6 7 8 98 日本語 ...
Page 112: ...그림 3 장착 예 1 흐름 통과 T 장착부 PVC 3 4 인치 NPT 2 흐름 통과 T 장착부 스테인리스 스틸 1 4 인치 NPT 112 한글 ...
Page 114: ...1 2 3 4 114 한글 ...
Page 115: ...5 6 7 8 한글 115 ...
Page 130: ...1 2 3 4 130 ไทย ...
Page 131: ...5 6 7 8 ไทย 131 ...
Page 142: ......
Page 143: ......