Figure 5 8315 Probe
Installation example for the 8394 probe
This probe installs perfectly in a Tri-Clover
®
Tri-Clamp
™
Tee starting from 1.5 inch diameter (
A
), and
also at a 90° angle starting from 2 inches (
B
). All Tri-Clamp
™
Tees are in compliance with 3A
standards for Cleaning In Place (CIP).
Figure 6 8394 Probe
In a bypass
POLYMETRON flow chambers are designed not to retain air bubbles. Minimum recommended flow
rate: 100 mL/minute (6 L/hour) with homogeneous sample flow or ideally 330 mL/minute (20 L/hour).
Note: The progressive accumulation of bubbles on the surface of the probe reduces the active surface, increases
the cell constant, and leads to an abnormally low conductivity measurement.