Specifications
Specifications are subject to change without notice.
Table 1 Probe specifications
8310 / 8315
8311 / 8316
Applications
Pure and ultra pure water
Moderately conductive solutions
K (cm
-1
)
0.01
0.1
Accuracy
< 2%
< 2%
Transmitter measurement range
0.01 to 200 μS.cm
-1
0.1 μS to 2 mS.cm
-1
Pt100 temperature response (t
90%)
< 30 seconds
< 45 seconds
Maximum temperature (°C)
125 (8310)
150 (8315)
125 (8311)
150 (8316)
Maximum pressure (bars)
10 (8310)
25 (8315)
10 (8311)
25 (8316)
Sample connection
¾ inch NPT
¾ inch NPT
8312 / 8317
8394
Applications
Waste water and drinking water
Food and pharmaceutical industries
(sterilized)
K (cm
-1
)
1.0
0.01
Accuracy
< 2%
< 2%
Transmitter measurement range
1 μS to 20 mS.cm
-1
0.01 to 200 μS.cm
-1
Pt100 temperature response (t
90%)
< 3 minutes
< 45 seconds
Maximum temperature (°C)
125 (8312)
150 (8317)
150
Maximum pressure (bars)
10 (8312)
25 (8317)
25
Sample connection
¾ inch NPT
Tri-Clamp 1½ or 2 inch
Table 2 Flow chamber specifications
08313=A=0001
08318=A=0001
Material
PVC
316 L SS
Maximum temperature (°C)
60 at 2 bar
150
Maximum pressure (bars)
15 at 25°C
25
Sensor connection
¾ inch NPT
¾ inch NPT
Sample connection
¾ inch NPT
¼ inch NPT
English
3