Note: For illustration purposes, cable part number 08319=A=00xx is shown in
Figure 4
.
1
External shielding
4
External electrode
2
Internal shielding
5
Pt100
3
Internal electrode
6
Pt100
The cable is available in lengths of 5, 10 or 20 meters and must be connected in compliance with the
following table:
Function
Color
External shielding
White (red tip)
Internal shielding
White (orange tip)
Internal electrode
White (yellow tip)
External electrode
Red
Pt100
Black
Pt100
Blue
Note: Refer to the user manual delivered with the transmitter for a detailed description of the cable connectors on
the transmitter.
Probe installation
In
Figure 5
on page 8,
Figure 6
on page 8 and
Figure 7
on page 9 the annotations
A
,
B
and
C
indicate:
•
A:
Ideal installation - perfect immersion of the electrode surfaces.
•
B:
Good installation - satisfactory immersion of the electrode surfaces.
•
C:
Poor installation - incomplete immersion of the electrodes, the conductivity will be too low.
On piping
Immerse the internal electrode completely in the process sample. For a 90° installation, take into
account the dimensions (see
Dimensions
on page 5).
Note: In the following illustrations, the arrows indicate the sample flow direction.
Installation example for the 8315 probe
English