![Hach ORBISPHERE 3650 Atex User Manual Download Page 13](http://html1.mh-extra.com/html/hach/orbisphere-3650-atex/orbisphere-3650-atex_user-manual_3336372013.webp)
11
Section 2
Specifications and Certifications
2.1 General technical data
2.2 Analyzer gas and display options
3650EX Instrument
Power Supply
Model 32960 non-rechargeable lithium battery
Battery Autonomy
60 hours continuous use
Signal Drift
< 0.5% of reading between service
Serial Output (RS232)
Baud rate: 9600; Stop Bits: 1; Start Bits: 0; Parity: None;
Temperature Compensation Range
-5 to 60°C
Instrument Operating Limits
0 to 45°C
Dimensions (HxWxD)
150 mm x 115 mm x 220 mm
Weight
2.4 kg
Enclosure
IP 65/NEMA 4
EMC Directive
EN 61326-1
ATEX Directive
EN 60079-0
EN 60079-11
LCIE 03 ATEX 6003 X
II 1 G, EX ia IIC T4 Ga
ISO Certification
ISO9001/EN29001
29122 Interface Box
Power Supply
120Vac 50/60Hz (Model 29122.A)
230Vac 50/60Hz (Model 29122.B)
Power Consumption
11VA (Model 29122.A)
7VA (Model 29122.B)
Fuse
Max current 250mA (Model 29122.A)
Max current 100mA (Model 29122.B)
Operating Limits
0 to 45°C
Dimensions (HxWxD)
70 mm x 140 mm x 190 mm
Weight
0.65 kg
Enclosure
IP 20
Enclosure material
ABS FR (V0)
EMC Directive
EN 61326-1
LVD Directive
EN 61010-1
Instrument Model
Gas Measured
Display Units
Maximum Display
Resolution
3650EX/111
Oxygen
ppm/ppb (liquid)
1 ppb
3650EX/112
Oxygen
%/ppm (gaseous)
1 ppm
3650EX/113
Oxygen
ppm (liquid)
% (gaseous)
0.001 ppm
0.001%
3650EX/114
Oxygen
kPa/Pa (gaseous)
1 Pa
3650EX/115
Oxygen
bar/mbar (gaseous)
1 mbar
3650EX/211
Hydrogen
ppm/ppb (liquid)
0.01 ppb
3650EX/212
Hydrogen
%/ppm (gaseous)
0.01 ppm