8
General Information
2.2 General sensor information
The sensor was developed for use in municipal waste water
applications.
The NO3D sc sensor (see
) with ion-selective electrode (ISE
sensor) is a continuous online process sensor for the measurement of
nitrate directly in the tank. It operates without reagents and requires no
further processing of the sample. The nitrate ions are measured using
an ion-selective electrode.
The only wear part is the sensor cartridge (see
)
(Catalog Number 6188401). This includes the ion-selective electrodes
for nitrate and chloride (compensation electrode), a pHD electrode
used as a reference system and a temperature sensor for temperature
compensation.
An optional cleaning unit is designed for automatically cleaning the
sensor head membranes and can be ordered separately. Refer to the
instruction sheet supplied with the cleaning unit.
The manufacturer recommends the use of the High Output Air Blast
system for the compressed air supply; this is a compressor in
weather-proof plastic housing
.
Figure 2 NO3D sc sensor
1
Sensor cartridge
3
Sensor adapter
2
Locking ring
4
Sensor housing
Figure 3 Sensor cartridge
1
Salt bridge
3
Temperature sensor
2
Membrane for chloride
4
Membrane for nitrate
Summary of Contents for NO3D sc
Page 2: ...2...
Page 4: ...4 Table of Contents Section 10 Certification 43 Appendix A Modbus Register 45 Index 47...
Page 6: ...6 Specifications 1 1 Dimensions Figure 1 Stainless steel sensor dimensions...
Page 10: ...10 General Information...
Page 16: ...16 Installation Figure 10 Example of sensor installation with Rail Mount Kit...
Page 18: ...18 Installation...
Page 28: ...28 Operation...
Page 38: ...38 Spare Parts and Accessories...
Page 44: ...44 Certification...
Page 48: ...48...