21
Operation
CALIBRATE(continued)
CAL.CONFIG.>
VALUE CORR
Value correction (see
)
Once value correction is complete, the correction data are displayed in the form of
the MATX2
1st NO3–N NO3DSC
Enter the NO
3
–N concentration (NO3D sc) of
the 1st point
1st CL- NO3DSC
Enter the Cl
-
concentration (NO3D sc) of the
1st point
1st NO3–N LAB
Enter the NO
3
–N concentration (lab) of the 1st
point
2nd NO3–N NO3DSC
Enter the NO
3
–N concentration (NO3D sc) of
the second point
2nd CL- NO3DSC
Enter the Cl
-
concentration (NO3D sc) of the
second point
2nd NO3–N LAB
Enter the NO
3
–N concentration (lab) of the
second point
-or-
CAL.CONFIG.>
PREVIOUS CAL
Selection of one of the last four matrix and value corrections performed
-or-
CAL.CONFIG.>
FACTORY CAL
Selection of the preset sensor code
DATE
Date of last calibration
SENSOR CODE
Enter the sensor code
CONFIGURE
EDIT NAME
Entry or editing of the name. Up to 10 alphanumeric characters.
MEAS UNITS
Selection of mg/L or ppm as measurement unit
SET PARAMETER
Selection of NO
3
–N or NO
3
TEMP UNITS
Selection of °C or °F as temperature unit.
TEMP CORR
Entry of the temperature correction (–1.5 to +1.5 °C or –2.7 to +2.7 °F)
RESPONSE TIME
Entry of the response time (30 sec to 300 sec)
LOG SETUP
Selects the interval for the data logging
(deactivate, 1, 2, 5, 10, 15, 30 min, 1, 2, 6 hr).
CL- COMPENSATE.
Selection of the chloride compensation: On, off or fixed value
SET DEFAULTS
Resets the configuration on the factory setting.
4.5 Sensor setup menu (continued)
Summary of Contents for NO3D sc
Page 2: ...2...
Page 4: ...4 Table of Contents Section 10 Certification 43 Appendix A Modbus Register 45 Index 47...
Page 6: ...6 Specifications 1 1 Dimensions Figure 1 Stainless steel sensor dimensions...
Page 10: ...10 General Information...
Page 16: ...16 Installation Figure 10 Example of sensor installation with Rail Mount Kit...
Page 18: ...18 Installation...
Page 28: ...28 Operation...
Page 38: ...38 Spare Parts and Accessories...
Page 44: ...44 Certification...
Page 48: ...48...