Option
Description
Location
Sets the location name to identify the Profibus network module. Default: network module
serial number
Status
Shows the condition of the Profibus network module.
• Status—Options: Please wait—Shows until the network module finds all of the
configured secondary devices or shows when the module configuration is new and
looks for devices connections. Ready—Shows when the network module is prepared to
send data to the Profibus. Online—Shows when the network module is connected to
the PLC and cyclic data is sent.
Note: Based on the number of devices, the duration of a new or initial telegram
configuration can take some minutes.
• PLC configuration error—Shows when the network card has received a wrong
configuration of a PLC (programmable logic controller), such as write to a read only
variable.
Input from PLC
• Datalog Interval—Sets the data log interval.
• Unit—For each text parameter (1 to 8), select a Parameter and Unit that will show on
the tag that will send the PLC as follows:
1.
Select Parameter and Unit.
2.
Select Search parameter and search for related parameters. For example, enter
DO to find all of the parameters related to dissolved oxygen. A message shows the
quantity of parameters found. Push OK.
3.
Select Parameter to go to the list of parameters found.
4.
Select one parameter, then push OK.
5.
Select Unit to select one unit based on the selected parameter.
6.
Apply changes to save the configuration.
7.
Do the steps again for the other text parameters or go back to the Profibus DP
menu.
Reset
configuration
Sets the configuration back to the factory settings.
4.
Push OK.
6.5.6 Configure the EtherNet/IP
The controller can connect to a PLC through Industrial Ethernet Protocol including a EtherNet/IP
solution. Ring and bus topologies are available.
Pre-requisites:
1.
Make sure that a EtherNet/IP module is installed in the controller. Refer to the documentation
supplied with the module.
2.
Make sure that all of the necessary electrical connections are complete before the module is
configured.
3.
Make sure to set the Ethernet ports option to IEP only or Mix IEP. Refer to
on page 25.
For configuration information, refer to the EtherNet/IP documentation, available on the
manufacturer's website.
6.5.7 Configure the PROFINET
The controller can connect to a PLC through the Industrial Ethernet Protocol, which includes a
PROFINET solution. Ring and bus topologies are available.
English
41