User Manual
DOC022.5
2
.80026
Refillable ORP/Redox Probe: Model MTC30101 or MTC30103
Safety information
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed. A symbol on the instrument is referenced in the
manual with a precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal
systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.
Specifications
Note: Specifications are subject to change without notice.
Specifications
Details
Probe type
Digital combination electrode with a refillable Ag/AgCl reference and a built-
in temperature sensor
Range
±1200 mV
Resolution
0.1 mV
Temperature accuracy
±0.3 °C (±0.54 °F)
Operating temperature range
0 to 80 °C (32 to 176 °F)
Storage temperature range
5 to 40 °C (41 to 104 °F)
Junction
Ceramic
Reference potential versus Standard
Hydrogen Electrode
207 mV at 25 °C
Fill solution
3M KCl with saturated AgCl
Reference type
Ag/AgCl (3 M KCl)
Minimum sample depth
20 mm
Dimensions
Diameter: 12 mm (0.47 in.)
Length: 175 mm (6.89 in.)
Cable length: 1 or 3 m (3.28 or 9.84 ft)
Cable connection
M12 digital output and connector compatible with HQd meters
Product overview
The MTC301 series probe is a refillable, combination oxidation reduction potential
(ORP/Redox) probe with a built-in temperature sensor (
Figure 1
). The MTC30101 or
MTC30103 probe is available with a 1 or 3 m (3.28 or 9.84 ft) cable and is intended for
laboratory use. The probe measures absolute mV values in wastewater, drinking water
and general applications. The probe measures ORP/Redox in wastewater, drinking water
and general applications. A 59 mL bottle of reference electrolyte filling solution (3M KCl
solution saturated with AgCl) is included with the probe.
1
Summary of Contents for Intellical MTC30101
Page 15: ...15...