background image

3.

Select Modify Current Settings.

4.

Select Measurement Options and update the settings:

Option

Description

Response Time

Sets the response time—

Fast (2 mV/minute)

Medium (1 mV/minute) (default)

Slow (0.5 mV/minute)

The response time affects the speed of the measurement by adjusting the
stabilization criteria.

Measurement
Limits

Sets the measurement limits—Lower limit (default: -1200.0 mV) or Upper
limit (default: 1200.0 mV).
The measurement limits can be set to match the acceptable values for
the sample. When the measurement is above the upper limit setting or
below the lower limit setting, the meter shows an "Out of limits" message.
This message is an alert to a potential problem with the process
conditions.

5.

If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.

6.

Push 

EXIT

 until the meter returns to the measurement mode.

Change calibration options

1.

Make sure a probe is connected to the meter.

2.

Push   and select MTC301 Settings.

3.

Select Modify Current Settings.

4.

Select Calibration Options and update the settings:

Option

Description

Standard

Sets the calibration standard—

ZoBell’s (221mV – 25 °C)

Light’s (468mV – 25 °C)

Custom

Temperature compensated for ZoBell’s solution.
Standard values are shown on the Calibration Options screen.
Light’s solution is characterized at 25 °C.
Custom standard values and temperature are user-defined.

Offset Limits

Sets the offset limits—±1 mV to 250 mV (default: ±25 mV).
The offset must fall within set limits for successful calibration.

Standard
Value

When Standard is set to Custom, sets the values for the custom calibration
standard— -1200.00 to 1200.0 mV (default: +221.0 mV). Custom standards
are characterized at 25 °C.

5.

Select Calibration Reminder and update the settings:

Option

Description

Reminder
Repeat

Meter will make an audible sound when a calibration is due and repeat the
sound at the selected interval—Off (default), 1 d, 7 d or 30 d.

Expires

Calibration expires after the selected time—Immediately, Re
30 min, Re 1 h, Re 2 h or Continue Reading.

Note: The meter cannot be used to read samples after calibration has

expired unless Continue Reading is selected.

9

Summary of Contents for Intellical MTC30101

Page 1: ...ng temperature range 0 to 80 C 32 to 176 F Storage temperature range 5 to 40 C 41 to 104 F Junction Ceramic Reference potential versus Standard Hydrogen Electrode 207 mV at 25 C Fill solution 3M KCl w...

Page 2: ...ode with deionized water thoroughly to remove the viscous storage filling solution completely Blot dry with a lint free cloth 4 Remove the protective tape from the filling hole before initial use refe...

Page 3: ...he temperature range of 0 to 30 C 32 to 86 F Light s solution should be read at 25 C 77 F Custom ORP Redox calibration solution values and temperature are user defined The calibration is recorded in t...

Page 4: ...er to the HQd meter manual for more information Regular calibration is required for the best measurement accuracy refer to Calibration on page 3 Prepare the probe for use refer to Preparation for use...

Page 5: ...ation record the sample pH and temperature 6 Repeat steps 3 6 for additional measurements 7 When measurements are done store the probe refer to Storage on page 11 Measurement conversion to SHE referen...

Page 6: ...ation state After each addition of titrant the ORP Redox electrode develops a potential proportional to the logarithm of the ratio of the activities of the two oxidation states At the inflection or en...

Page 7: ...normality of the sample Nx in equivalents per liter Nx Vt x Nt Vx where Nt normality of titrant Eq L Vt volume of titrant at end point mL Vx Volume of sample mL Run Check Standard The Run Check Stand...

Page 8: ...available options and how to change them are given in the screens tables and procedures throughout this section The settings that can be changed are shown in Table 2 Table 2 Parameter specific settin...

Page 9: ...ush and select MTC301 Settings 3 Select Modify Current Settings 4 Select Calibration Options and update the settings Option Description Standard Sets the calibration standard ZoBell s 221mV 25 C Light...

Page 10: ...stom check standard are user defined 5 Select Check Standard Reminder and update the settings Option Description Reminder Repeat Sets the time interval for the check standard reminder Off default 1 d...

Page 11: ...uct overview on page 1 2 Remove the cap from the tip of the filling solution bottle 3 Hold the bottle so that the tip is down Put the tip of the bottle in the filling hole 4 Slowly squeeze the bottle...

Page 12: ...r manual Large number of methods stored on the probe Continue to let the probe connect Do not disconnect the probe Damaged probe Make sure there is connectivity with another probe or meter to confirm...

Page 13: ...onse time Platinum electrode not conditioned for reducing type samples 1 Make sure that the platinum disc is clean and smooth 2 Put the platinum disc in Reducing Solution for ORP Electrodes for 3 10 m...

Page 14: ...e probe downward in the solution sample to remove any air bubbles in the reference junction holes Out of limits Check standard value is outside of limits set in the current method Make sure that the s...

Page 15: ...15...

Page 16: ...ch com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info hach lange de www hach lange de HACH LANGE S rl 6 route de Compois 1222 V se...

Reviews: