6.
Push
Read
.
Stir gently. The
display will show
"Stabilizing" and a
progress bar as
the probe
stabilizes in the
sample. The
display shows the
standard solution
value and the mV
offset when the
reading is stable.
7.
Push
Done
to
view the
calibration
summary.
8.
Push
Store
to
accept the
calibration and go
back to the
measurement
mode.
Sample measurement
Before measurement:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
If complete traceability is necessary, enter a sample ID and operator ID before measurement. Refer to the HQd meter
manual for more information.
Regular calibration is required for the best measurement accuracy (refer to
Calibration
on page 3).
Prepare the probe for use (refer to
Preparation for use
on page 2).
Make sure that the platinum disc is clean and smooth (refer to
Clean the probe
on page 10).
Measurement notes:
•
Data is automatically stored in the data log when
Press to Read
or
Interval
is
selected in the Measurement Mode. When
Continuous
is selected, data will only be
stored when
Store
is selected.
•
Air bubbles under the sensor tip when submerged can cause slow response or error
in measurement. If bubbles are present, gently shake the probe until bubbles are
removed.
•
If a measurement error occurs, refer to
Troubleshooting
on page 12.
Measurement—direct method procedure:
4