Problem
Possible cause
Solution
The readings
are low.
The wrong parameter
was selected.
Make sure to select the correct
parameter (e.g., free chlorine) and
use the correct reagents for the
parameter.
The readings
are low.
The reagents have
expired.
Discard the set of reagent bottles.
Use a new set of reagent bottles. Do
a reagent blank calibration. Refer to
Do a reagent blank calibration
on page 16.
The readings
are high.
The reagent blank
value is not correct or
missing.
Make sure to do a reagent blank
calibration before a set of new
reagent bottles is used. Refer to
on page 16.
If the problem continues, do a
calibration check. Refer to
The readings
are high.
The sample vials are
dirty or have scratches.
Clean the sample vials if dirty.
Replace the sample vials if the
sample vials have scratches.
The readings
are high.
The wrong parameter
was selected.
Make sure to select the correct
parameter (e.g., free chlorine) and
use the correct reagents for the
parameter.
Unexpected
readings
The sample is not a
clean water sample.
The DR1300 FL is intended for use
with clean water samples.
24
English
Summary of Contents for DR1300 FL
Page 2: ...Table of Contents English 3 Espa ol 27 Fran ais 54 Portugu s 81 108 129 153 2...
Page 112: ...1 DR1300 FL 1 3 3 2 4 LCD 3 112...
Page 113: ...3 4 2 2 1 DR1300 FL 4 2 5 USB 4 3 2 6 AA 4 113...
Page 114: ...4 AA 3 3 114...
Page 115: ...5 5 1 4 4 1 Enter 5 4 Right 2 Left 5 Down 3 Up 115...
Page 116: ...5 2 5 5 1 5 2 6 3 7 4 116...
Page 118: ...7 MSDS SDS 119 1 Enter 2 3 a b c d 119 e f g 4 118...
Page 124: ...10 10 1 AA 114 10 2 10 3 124...
Page 125: ...11 120 120 120 120 125...
Page 126: ...124 120 120 120 DR1300 FL DPD ppb 12 126...
Page 128: ...1 3563500 16 mm 6 pkg 100866 128...
Page 131: ...3 1 3 1 1 131...
Page 133: ...1 2 3 3 2 Hach DR1300 FL DR1300 FL DR1300 FL DR1300 FL 3 3 2 DR1300 FL 1 DR1300 FL 133...
Page 134: ...1 DR1300 FL 1 3 3 2 4 LCD 3 134...
Page 135: ...3 4 2 2 1 DR1300 FL 4 2 5 Bluetooth USB 4 3 2 6 AA 4 4 Bluetooth 135...
Page 136: ...4 AA 4 3 3 136...
Page 137: ...5 5 1 4 4 1 Enter 5 4 2 5 3 137...
Page 138: ...5 2 5 5 1 SYSTEM 5 2 TESTS 6 CALIBRATION 3 7 POWER 4 138...
Page 140: ...Update Program Factory Default Factory Exit 7 MSDS SDS 141 1 Enter 2 TESTS 140...
Page 141: ...3 a b TIMER START c d 141 e f g BACK READ 4 7 1 7 2 1 http www hach com 2 DR1300 FL 3 141...
Page 147: ...LOW BATTERY 4 AA 136 10 2 10 3 147...
Page 148: ...11 SYSTEM Diagnosis Sample Collection 142 143 148...
Page 149: ...142 142 147 142 142 143 149...
Page 150: ...DR1300 FL Sample Collection DPD ppb 12 150...
Page 152: ...1 3563500 16mm 6 100866 152...
Page 155: ...3 1 1 3 1 2 3 1 3 ICES 003 155...
Page 157: ...DR1300 FL 1 DR1300 FL 1 DR1300 FL 1 3 3 Keypad 2 4 LCD 3 157...
Page 158: ...3 4 2 2 1 DR1300 FL 4 2 5 Bluetooth USB4 3 2x 6 AA 4 Bluetooth 158...
Page 159: ...4 AA 3 3 159...
Page 160: ...5 5 1 Keypad 4 4 1 Enter 5 4 2 5 3 160...
Page 161: ...5 2 Display 5 5 Main menu 1 SYSTEM 5 2 TESTS 6 CALIBRATION 3 7 POWER 4 161...
Page 163: ...Factory Default Factory Exit 7 MSDS SDS 164 1 Enter 2 TESTS 3 a 163...
Page 165: ...8 MSDS SDS 8 1 1 Enter 2 3 164 4 CALIBRATION 5 CAL RB 6 a b c d e Enter 7 Enter 8 165...
Page 170: ...11 SYSTEM Diagnosis 165 166 165 170...
Page 171: ...165 169 165 165 166 DR1300 FL DPD 171...
Page 174: ......
Page 175: ......