Section 11 Troubleshooting
If an error occurs, a warning message will show on the display with
instructions.
If the instrument does not respond, remove the batteries and then
install the batteries. Replace the batteries if the battery power level is
low.
To do a diagnostic test, select
SYSTEM
>
Diagnosis
with the sample
port cover closed. Supply the diagnostic data to technical support for
analysis.
Problem
Possible cause
Solution
The instrument
will not come
on.
The batteries are not
installed in the correct
orientation.
The battery power level
is low.
Make sure the battery orientation is
correct.
Replace the batteries if the battery
power level is low. Use only non-
rechargeable batteries.
The chlorine
readings are
low.
The sample containers,
sample vials, caps and
other items that
touched the sample
have a chlorine
demand.
Collect samples in clean glass
bottles. Do not use plastic
containers. Plastic containers can
have a large chlorine demand.
Pretreat items that the sample may
touch to remove chlorine demand.
Refer to the instructions in the
Sample Collection
section of the test
procedure.
The chlorine
readings are
low.
The sample was not
analyzed immediately.
Analyze the samples immediately.
Chlorine is a strong oxidizing agent
and is unstable in natural waters.
The total
chlorine
readings are
high or low.
The reagents were
added in an incorrect
sequence.
Make sure that the reagents are
added to the sample vial in the
sequence given in the test
procedure.
22
English
Summary of Contents for DR1300 FL
Page 2: ...Table of Contents English 3 Espa ol 27 Fran ais 54 Portugu s 81 108 129 153 2...
Page 112: ...1 DR1300 FL 1 3 3 2 4 LCD 3 112...
Page 113: ...3 4 2 2 1 DR1300 FL 4 2 5 USB 4 3 2 6 AA 4 113...
Page 114: ...4 AA 3 3 114...
Page 115: ...5 5 1 4 4 1 Enter 5 4 Right 2 Left 5 Down 3 Up 115...
Page 116: ...5 2 5 5 1 5 2 6 3 7 4 116...
Page 118: ...7 MSDS SDS 119 1 Enter 2 3 a b c d 119 e f g 4 118...
Page 124: ...10 10 1 AA 114 10 2 10 3 124...
Page 125: ...11 120 120 120 120 125...
Page 126: ...124 120 120 120 DR1300 FL DPD ppb 12 126...
Page 128: ...1 3563500 16 mm 6 pkg 100866 128...
Page 131: ...3 1 3 1 1 131...
Page 133: ...1 2 3 3 2 Hach DR1300 FL DR1300 FL DR1300 FL DR1300 FL 3 3 2 DR1300 FL 1 DR1300 FL 133...
Page 134: ...1 DR1300 FL 1 3 3 2 4 LCD 3 134...
Page 135: ...3 4 2 2 1 DR1300 FL 4 2 5 Bluetooth USB 4 3 2 6 AA 4 4 Bluetooth 135...
Page 136: ...4 AA 4 3 3 136...
Page 137: ...5 5 1 4 4 1 Enter 5 4 2 5 3 137...
Page 138: ...5 2 5 5 1 SYSTEM 5 2 TESTS 6 CALIBRATION 3 7 POWER 4 138...
Page 140: ...Update Program Factory Default Factory Exit 7 MSDS SDS 141 1 Enter 2 TESTS 140...
Page 141: ...3 a b TIMER START c d 141 e f g BACK READ 4 7 1 7 2 1 http www hach com 2 DR1300 FL 3 141...
Page 147: ...LOW BATTERY 4 AA 136 10 2 10 3 147...
Page 148: ...11 SYSTEM Diagnosis Sample Collection 142 143 148...
Page 149: ...142 142 147 142 142 143 149...
Page 150: ...DR1300 FL Sample Collection DPD ppb 12 150...
Page 152: ...1 3563500 16mm 6 100866 152...
Page 155: ...3 1 1 3 1 2 3 1 3 ICES 003 155...
Page 157: ...DR1300 FL 1 DR1300 FL 1 DR1300 FL 1 3 3 Keypad 2 4 LCD 3 157...
Page 158: ...3 4 2 2 1 DR1300 FL 4 2 5 Bluetooth USB4 3 2x 6 AA 4 Bluetooth 158...
Page 159: ...4 AA 3 3 159...
Page 160: ...5 5 1 Keypad 4 4 1 Enter 5 4 2 5 3 160...
Page 161: ...5 2 Display 5 5 Main menu 1 SYSTEM 5 2 TESTS 6 CALIBRATION 3 7 POWER 4 161...
Page 163: ...Factory Default Factory Exit 7 MSDS SDS 164 1 Enter 2 TESTS 3 a 163...
Page 165: ...8 MSDS SDS 8 1 1 Enter 2 3 164 4 CALIBRATION 5 CAL RB 6 a b c d e Enter 7 Enter 8 165...
Page 170: ...11 SYSTEM Diagnosis 165 166 165 170...
Page 171: ...165 169 165 165 166 DR1300 FL DPD 171...
Page 174: ......
Page 175: ......