4.
Scroll down to "Methods/Procedures".
5.
Click the link for the applicable test procedure to download it.
Section 8 Calibration
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and
wear all of the personal protective equipment appropriate to the
chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
8.1 Do a reagent blank calibration
Do a reagent blank calibration before a set of new reagent bottles is
used.
1.
Push
Enter
to set the instrument to on. Keep the sample port cover
closed during startup.
2.
After startup, make sure that the sample port is clean. Use a soft
cloth or no-lint paper tissue (or air duster) to clean the sample port
as necessary.
3.
Install the vial adapter in the instrument. Push in the tabs on the
sides of the vial adapter. Refer to the illustrated steps in
vial adapter, vial and vial cover
4.
Select
CALIBRATION
, then select the parameter for the
calibration.
5.
Select
CAL-RB
.
6.
Measure a reagent blank as follows:
a.
Prepare a reagent blank. Replace the sample with deionized
water (or distilled water or organic-free water) in the test
procedure. Do not use the timer.
b.
Clean the external surface of the sample vial with a no-lint cloth.
c.
Install the sample vial in the vial adapter.
d.
Put the vial cover on the sample vial.
e.
Push
Enter
to measure the reagent blank.
16
English
Summary of Contents for DR1300 FL
Page 2: ...Table of Contents English 3 Espa ol 27 Fran ais 54 Portugu s 81 108 129 153 2...
Page 112: ...1 DR1300 FL 1 3 3 2 4 LCD 3 112...
Page 113: ...3 4 2 2 1 DR1300 FL 4 2 5 USB 4 3 2 6 AA 4 113...
Page 114: ...4 AA 3 3 114...
Page 115: ...5 5 1 4 4 1 Enter 5 4 Right 2 Left 5 Down 3 Up 115...
Page 116: ...5 2 5 5 1 5 2 6 3 7 4 116...
Page 118: ...7 MSDS SDS 119 1 Enter 2 3 a b c d 119 e f g 4 118...
Page 124: ...10 10 1 AA 114 10 2 10 3 124...
Page 125: ...11 120 120 120 120 125...
Page 126: ...124 120 120 120 DR1300 FL DPD ppb 12 126...
Page 128: ...1 3563500 16 mm 6 pkg 100866 128...
Page 131: ...3 1 3 1 1 131...
Page 133: ...1 2 3 3 2 Hach DR1300 FL DR1300 FL DR1300 FL DR1300 FL 3 3 2 DR1300 FL 1 DR1300 FL 133...
Page 134: ...1 DR1300 FL 1 3 3 2 4 LCD 3 134...
Page 135: ...3 4 2 2 1 DR1300 FL 4 2 5 Bluetooth USB 4 3 2 6 AA 4 4 Bluetooth 135...
Page 136: ...4 AA 4 3 3 136...
Page 137: ...5 5 1 4 4 1 Enter 5 4 2 5 3 137...
Page 138: ...5 2 5 5 1 SYSTEM 5 2 TESTS 6 CALIBRATION 3 7 POWER 4 138...
Page 140: ...Update Program Factory Default Factory Exit 7 MSDS SDS 141 1 Enter 2 TESTS 140...
Page 141: ...3 a b TIMER START c d 141 e f g BACK READ 4 7 1 7 2 1 http www hach com 2 DR1300 FL 3 141...
Page 147: ...LOW BATTERY 4 AA 136 10 2 10 3 147...
Page 148: ...11 SYSTEM Diagnosis Sample Collection 142 143 148...
Page 149: ...142 142 147 142 142 143 149...
Page 150: ...DR1300 FL Sample Collection DPD ppb 12 150...
Page 152: ...1 3563500 16mm 6 100866 152...
Page 155: ...3 1 1 3 1 2 3 1 3 ICES 003 155...
Page 157: ...DR1300 FL 1 DR1300 FL 1 DR1300 FL 1 3 3 Keypad 2 4 LCD 3 157...
Page 158: ...3 4 2 2 1 DR1300 FL 4 2 5 Bluetooth USB4 3 2x 6 AA 4 Bluetooth 158...
Page 159: ...4 AA 3 3 159...
Page 160: ...5 5 1 Keypad 4 4 1 Enter 5 4 2 5 3 160...
Page 161: ...5 2 Display 5 5 Main menu 1 SYSTEM 5 2 TESTS 6 CALIBRATION 3 7 POWER 4 161...
Page 163: ...Factory Default Factory Exit 7 MSDS SDS 164 1 Enter 2 TESTS 3 a 163...
Page 165: ...8 MSDS SDS 8 1 1 Enter 2 3 164 4 CALIBRATION 5 CAL RB 6 a b c d e Enter 7 Enter 8 165...
Page 170: ...11 SYSTEM Diagnosis 165 166 165 170...
Page 171: ...165 169 165 165 166 DR1300 FL DPD 171...
Page 174: ......
Page 175: ......