Spec Filter accuracy testing—DR 5000 and DR 2800

5.

Enter the maximum wavelength absorbance into the test record in the "Actual 
value" column for the Ho filter. Open the lid and remove the filter.

6.

Select

OPTIONS / 

λ 

and enter 806–809, step 0.1 nm, then press 

OK

.

7.

Press

ZERO

.

8.

Insert the BG20/2 filter into the 10-mm square cell holder and close the lid. 
Press

READ

.

9.

Enter the maximum wavelength absorbance into the test record in the "Actual 
value" column for the BG20/2 filter. Open the lid and remove the filter.

Wavelength accuracy—DR 2800

1.

Install the 10-mm square cell holder (Adapter A). Turn on the instrument.

2.

Select

SYSTEM CHECKS>OPTICAL CHECKS>WAVELENGTH CHECK

.

3.

Insert the BG20/2 filter into the 10-mm square cell holder and close the lid. 
Press

START

.

4.

Enter the maximum wavelength absorbance into the test record in the “Actual 
value” column for the BG20/2 filter. Open the lid and remove the filter.

5.

Press

CANCEL

.