2

Spec Filter accuracy testing—DR 5000 and DR 2800

Test Procedures

Stray light test

1.

Install the 10-mm square cell holder (Adapter A with the DR 2800). Turn on 
the instrument.

2.

Press 

SINGLE WAVELENGTH

, then 

OPTIONS 

/

λ

.

3.

Enter test 

340

 and press 

OK

.

4.

Press 

ZERO

.

5.

Insert the KV450/3 filter into the 10-mm square cell holder and close the lid. 
Press

READ

.

6.

Enter the measured value into the test record under “Actual value” for the KV450/3 
filter. Open the lid and remove the filter.

Photometric accuracy

1.

Install the 10-mm square cell holder (Adapter A with the DR 2800). Turn on 
the instrument.

2.

Press 

OPTIONS / 

λ

, enter 

546

, then press 

OK

.

3.

Press 

ZERO

.

4.

Insert NG9/1 filter into the square cell holder then close the lid. Press 

READ

.

5.

Enter the measured value into the test record in the “Actual value” column for the 
NG9/1 filter. Open the lid and remove the filter.

6.

Insert the NG5/2 filter into the 10-mm square cell holder then close the lid. 
Press

READ

.

7.

Enter the measured value into the test record in the "Actual value" column for the 
NG5/2 filter. Open the lid and remove the filter.

8.

Insert the NG11/2 filter into the 10-mm square cell holder then close the lid. 
Press

READ

.

9.

Enter the measured value into the test record in the "Actual value" column for the 
NG11/2 filter. Open the lid and remove the filter.

10.

Press 

EXIT

 or select 

Main Menu

.

Wavelength accuracy—DR 5000

1.

Install the 10-mm square cell holder. Turn on the instrument.

2.

Press 

SINGLE WAVELENGTH

, then 

OPTIONS / 

λ

. Enter 358–362, step 0.1 nm, then 

press 

OK

.

3.

Press 

ZERO

.

4.

Insert the Ho filter into the 10-mm square cell holder and close the lid. Press 

READ

.