Use the external trigger input to start a measurement. The trigger input can be operated with a
12–24 V DC (± 5%) voltage. When the trigger is set to off, a measurement is started for all of the
sensors that have automatic measurement enabled.
Note: The manufacturer recommends to disable the automatic measurement when an external trigger is used.
1.
Remove the power to the instrument.
2.
Open the controller cover. Refer to
on page 10.
3.
Put the cable through the relay strain relief fitting. Refer to
Electrical connectors and fittings
on page 9.
4.
Prepare the wires. Refer to the wiring requirements in
on page 3.
5.
Pull to remove the connector.
6.
Install each wire in the connector. Refer to
.
7.
Tighten the cable strain relief.
8.
Close the controller cover.
Figure 8 Analog outputs
Table 6 Wiring information—Analog outputs and external trigger
CON9 Terminal
Description
CON 10 Terminal
Description
1
Output 1 +
1
Output 5 +
2
Output 1 –
2
Output 5 –
3
Output 2 +
3
Output 6 +
4
Output 2 –
4
Output 6 –
5
Output 3 +
5
Trigger input +
6
Output 3 –
6
Trigger input –
7
Output 4 +
7
Not used
8
Output 4 –
8
English
15