To start a test,
press the channel
number corresponding
to the bottle.
Note: Each channel
(1–6) must be started
individually.
14.
Press the
ON
key.
A menu for selecting
the BOD range will
be displayed.
15.
For 0–350 mg/L
range, press the
>
(right) key. For 0–700
mg/L press the
>
key a
second time.
For 0–35 mg/L range,
press the
<
(left) key.
For 0–70 mg/L press
the
<
key a second time.
BODTrak Test Procedure,
continued