Section 6
Reagents and Calibration Standards
6.1 Reagents
BioTector TOC analyzer uses the following reagents:
I.
Acid:
6.0 N Sulfuric Acid (H
2
SO
4
) Reagent containing 350 mg/l Manganese Sulfate Monohydrate
II.
Base:
4.0 N Sodium Hydroxide (NaOH) Reagent
Reagents should not contain high levels of organics, nitrates and phosphates. Ideally, the level of organics,
nitrates and phosphates should be less than 100
g/l (ppb) in the deionized water used to prepare TOC
analyzer reagents.
Acid and Base reagents are stable up to 1 year. The reagents must be stored in a safe and secure location,
where the temperature does not drop below 2°C, in compliance with the local regulations. Table 6 below
summarizes the total days each BioTector TOC Analyzer reagent lasts at various system configurations:
Table 6 BioTector TOC Analyzer Reagent Consumption
REAGENTS
Container Size
(Liters)
TOTAL DAYS REAGENT LASTS
Acid
19
239
Base
19
239
Above table is derived from several online operation parameters such as 100% online time.
Recommended bunds (reagent spill trays) to contain above quantity reagents are 1x 50 Liters.
Summary of Contents for BioTector B3500s
Page 17: ...Page 17 Software Menu Diagram...
Page 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Page 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Page 63: ...Page 63...
Page 74: ...Page 74 Section 8 Maintenance Menu Maintenance Menu Diagram...
Page 156: ...Page 156 Section 11 System Replacement and Spare Parts...
Page 162: ...Page 162 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...
Page 165: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...