95_LO AIR PRESSURE 1
Air pressure supply has been less than
0.8 bar for longer than 60 seconds.
When the oxygen concentrator is
running, the pressure typically cycles
from 1.5 bar to 0.9 bar.
Extreme fluctuations in air supply
pressure or faulty external air
regulator.
Increase air supply pressure to
1.5 bar when the oxygen
concentrator is not running.
When the problem is rectified,
acknowledge the warning by
pressing the ENTER key in the
Fault Archive menu, to reset the
O2-Controller Board.
96_HI O2 PRESSURE 2
Oxygen pressure supply has been
more than 500 mbar for longer than 5
seconds.
If system does not detect the oxygen
pressure drop to normal levels, the air
supply is isolated from the system and
oxygen is not generated.
Reduce oxygen pressure to 400
mbar (±10mbar) at 20 l/h MFC
flow, looking at the Oxygen
Controller Status (O2-CTRL
STATUS) menu and using the
oxygen pressure regulator.
When the problem is rectified,
acknowledge the warning by
pressing the ENTER key in the
Fault Archive menu, to reset the
O2-Controller Board.
97_HI O2 PRESSURE 1
Oxygen pressure supply has been
more than 450 mbar for longer than 60
seconds.
Reduce oxygen pressure to 400
mbar (±10mbar) at 20 l/h MFC
flow, looking at the Oxygen
Controller Status (O2-CTRL
STATUS) menu and using the
oxygen pressure regulator.
98_LO O2 PRESSURE 2
Oxygen pressure supply has been less
than 150 mbar for longer than 5
seconds.
If system does not detect the oxygen
pressure increase to normal levels, the
air supply is isolated from the system
and oxygen is not generated.
Increase oxygen pressure to 400
mbar (±10mbar) at 20 l/h MFC
flow, looking at the Oxygen
Controller Status (O2-CTRL
STATUS) menu and using the
oxygen pressure regulator.
When the problem is rectified,
acknowledge the warning by
pressing the ENTER key in the
Fault Archive menu, to reset the
O2-Controller Board.
99_LO O2 PRESSURE 1
Oxygen pressure supply has been less
than 200 mbar for longer than 60
seconds.
Increase oxygen pressure to 400
mbar (±10mbar) at 20 l/h MFC
flow, looking at the Oxygen
Controller Status (O2-CTRL
STATUS) menu and using the
oxygen pressure regulator.
Summary of Contents for BioTector B3500s
Page 17: ...Page 17 Software Menu Diagram...
Page 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Page 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Page 63: ...Page 63...
Page 74: ...Page 74 Section 8 Maintenance Menu Maintenance Menu Diagram...
Page 156: ...Page 156 Section 11 System Replacement and Spare Parts...
Page 162: ...Page 162 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...
Page 165: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...