As can be seen in the figure 13 above, base (Sodium Hydroxide) reagent container does not contain any vent
(breathing) hole. The breathing air into the base container is supplied through the CO
2
filter, which must be
fitted on the base container lid.
Figure 14 below shows the detailed connections on BioTector base reagent. The purpose of the CO
2
filter is
to prevent the base reagent coming in contact with atmospheric CO
2
present in the air. The soda lime inside
the CO
2
filter absorbs the atmospheric CO
2
and prevents the base reagent getting contaminated. If any vent
hole is accidentally drilled on the lid of the base container and if the fittings are not connected correctly on the
base reagent, contamination will occur and the background CO
2
readings will increase.
Figure 14 BioTector Sodium Hydroxide reagent dip tube setup
The vent port on the acid reagent dip tube should not be isolated or blocked. Failure to do so may cause the
container to collapse and leak.
The length of the dip tubes in all reagent containers should be adjusted correctly for the optimum usage of the
reagents. If HCl acid is used, stainless steel (SS-316) weights/filters should not be used in the reagent. The
recommended weights/filters for such reagents are PFA.
Summary of Contents for BioTector B3500s
Page 17: ...Page 17 Software Menu Diagram...
Page 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Page 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Page 63: ...Page 63...
Page 74: ...Page 74 Section 8 Maintenance Menu Maintenance Menu Diagram...
Page 156: ...Page 156 Section 11 System Replacement and Spare Parts...
Page 162: ...Page 162 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...
Page 165: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...