Page 141
9.2 BioTector Warning Event Explanation and Remedial Action
The following warning conditions do not activate the BioTector stop sequence and leave the 4-20mA signals
un-changed and do not activate the fault relay. When the system warning is rectified, acknowledge the
warning by pressing the ENTER key in the Fault Archive menu.
WARNING
CONDITION
CAUSE/REMEDY
21_CO2 ANL LENS DIRTY
CO
2
Analyzer has detected dirt on its
optical system.
Clean the CO
2
Analyzer.
Clean the lenses on the CO
2
Analyzer.
22_FLOW WARNING
– EX
MFC oxygen flow through the Exhaust
“EX” Valve (MV1) has dropped below
the “FLOW WARNING” level defined in
Pressure/Flow Test Program menu
8.3.4.5 Pressure/Flow Test Program
during the Pressure/Flow Test cycle.
Various, for example,
air and oxygen supply problem.
Check oxygen pressure looking at
the Oxygen Controller Status (O2-
CTRL STATUS) menu. It should
be 400mbar (±10mbar) at 20 l/h
MFC flow.
Partially blocked ozone destructor.
Partially blocked tube after MFC.
Faulty/blocked Exhaust Valve.
Faulty MFC.
Run Flow Test (see Section
23_FLOW WARNING
– SO
MFC oxygen flow through the Sample
Out
“SO” port (through Reactor Valve,
MV3)
has dropped below the “FLOW
W
ARNING” level defined in
Pressure/Flow Test Program menu
8.3.4.5 Pressure/Flow Test Program
during the Pressure/Flow Test cycle.
Various, for example,
air and oxygen supply problem.
Check oxygen pressure looking at
the Oxygen Controller Status (O2-
CTRL STATUS) menu. It should
be 400mbar (±10mbar) at 20 l/h
MFC flow.
Faulty/blocked Reactor Valve
and/or Sample Valve.
Partially blocked tube after MFC.
Faulty MFC.
Run Flow Test (see Section
26_PRESSURE TEST
WARN
MFC flow did not drop below the
“PRESSURE TEST WARNING” level
defined in Pressure/Flow Test Program
menu
cycle.
Small gas/liquid leak in the
BioTector.
Valve leaking or not sealing.
Open Reactor Valve, inspect for
dirt/damage.
Check system fittings.
Check Mixer Reactor.
Run Pressure Test (see Section
28_NO PRESSURE TEST
This warning is logged when the
pressure test procedure is skipped
during the system startup sequence.
The warning cannot be acknowledged
by the user. It can only be
acknowledged automatically by the
system next time the pressure test is
passed.
This warning occurs when the
Pressure/Flow Test is deactivated,
and when the “QUICK STARTUP”
function is used. See Section
Summary of Contents for BioTector B3500s
Page 17: ...Page 17 Software Menu Diagram...
Page 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Page 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Page 63: ...Page 63...
Page 74: ...Page 74 Section 8 Maintenance Menu Maintenance Menu Diagram...
Page 156: ...Page 156 Section 11 System Replacement and Spare Parts...
Page 162: ...Page 162 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...
Page 165: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...