SYSTEM PROGRAM 2 TC
SAMPLE IN 5p , P-V
▪
SAMPLE IN defines the amount of sample (in pulses) to be
injected into the reactor in TC mode at range 2.
SYSTEM PROGRAM 2 VOC
▪
All system parameters for SYSTEM PROGRAM 2 VOC analysis
mode are covered in System Program 2 TC and System
Program 2 TIC + TOC above.
SYSTEM PROGRAM 3
SYSTEM PROGRAM 3 TIC + TOC
RANGE CHANGE 2-3 150.0 mgC/l
▪
RANGE CHANGE 2-3 defines the concentration level for the
BioTector to change its range automatically from range 2 to
range 3.
▪
The RANGE CHANGE 2-3 is typically the top concentration
point of TOC RANGE 2 in System Range Data menu by default.
RANGE CHANGE 3-2 120.0mgC/l
▪
RANGE CHANGE 3-2 specifies the concentration level for the
BioTector to change its range automatically from range 3 to
range 2.
▪
The RANGE CHANGE 3-2 is typically 20% less of the top
concentration point of TOC RANGE 2 in System Range Data
menu by default.
RANGE CHANGE 3-1 40.0mgC/l
▪
RANGE CHANGE 3-1 defines the concentration level for the
BioTector to change its range automatically from range 3 to
range 1.
▪
The RANGE CHANGE 3-1 is typically 20% less of the top
concentration point of TOC RANGE 1 in System Range Data
menu by default.
SAMPLE IN 3p P-V
▪
SAMPLE IN states the amount of sample (in pulses) to be
injected into the reactor at range 3.
TIC ACID 12p
▪
TIC ACID defines the amount of acid reagent (in pulses) to be
injected into the reactor at range 3.
BASE 70p
▪
BASE specifies the amount of base (in pulses) to be injected into
the reactor for the BASE OXIDATION phase at range 3.
BASE OXIDATION 130s
▪
BASE OXIDATION defines the oxidation time (150s by default),
which can be programmed independently for range 3.
▪
If the Base Oxidation time is different than the one programmed
in System Program 1 menu, depending on the difference
between the two settings, the reaction time at range 3 will be
either longer or shorter than the reaction time displayed in
System Program 1 menu.
TOC ACID 58p
▪
TOC ACID states the amount of acid (in pulses) to be injected
into the reactor for the TOC SPARGE phase at range 3.
Summary of Contents for BioTector B3500s
Page 17: ...Page 17 Software Menu Diagram...
Page 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Page 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Page 63: ...Page 63...
Page 74: ...Page 74 Section 8 Maintenance Menu Maintenance Menu Diagram...
Page 156: ...Page 156 Section 11 System Replacement and Spare Parts...
Page 162: ...Page 162 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...
Page 165: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...