7.
Set the options as follows.
Option
Set to:
Delay Type
Date and Time
Date
Enter the start date for the sampling program.
Time
Enter the start time for the sampling program.
8.
Go directly to
Select when the program stops
on page 20.
Select when the sampling program starts
Select when the sampling program starts after the RUN/HALT button is pushed.
1.
Push
MENU
.
2.
Go to Programming>Sample Programming>Program Start.
3.
Select an option.
Option
Description
Immediately on
'Run'
The sampling program starts when
RUN/HALT
is pushed.
On Trigger
The sampling program starts (or starts and stops) when a trigger occurs. The trigger is a
selected channel alarm or a signal from an external flow meter or third-party device that
is connected to the AUX I/O port. Refer to
Start the program on a trigger (optional)
on page 20.
After Delay
When
RUN/HALT
is pushed, the sampling program starts after the selected delay.
Refer to
Table 11
for the delay options.
On Schedule
The sampling program starts on the selected day(s) of the week and time(s). Select a
maximum of 12 schedule intervals.
Delay
—Sets the sampling program to start on the first scheduled day and time after a
selected delay. The delay is one time only. Refer to
Table 11
for the delay options.
Schedule Intervals
—Sets the number of schedule intervals (1 to 12).
Edit Schedule
—
Sets the start day and time and the stop day and time for each schedule interval.
Table 11 Delay options
Type
Description
None
The sampling program starts after no delay.
Data and Time
The sampling program starts after a set date and time.
Time Duration
The sampling program starts after a set amount of time.
Flow Duration
The sampling program starts after a set flow volume occurs. Select the source of the flow
measurement.
AUX-Pulse (or AUX-mA)
—An external flow meter or third-party device connected to the
AUX I/O port.
Flow 1 (or 2)
—A sensor connected to the Sensor 1 (or 2) port.
IO 1 (or IO 2)
—Analog Input IO 1 (or IO 2) of the optional IO9000 module.
Note: A flow measurement that is supplied through a sensor port or the IO9000 module is only available for
selection after it is selected to be recorded to the data log.
Slave, Sequential
4
The sampling program starts after the sampling program of the leading sampler is complete.
Slave, Parallel
4
The sampling program operates only when the sampling program of the leading sampler
operates.
4
This option is only applicable when two sampling programs are used. If a slave delay setting is
not selected, the two sampling programs operate independently.
English
19