Table 7 Primary device options
Option
Description
Area Velocity Select the method of calculating the area, then go to
Table 8
. Options: Circular, Rectangular,
Trapezoidal or U-Shaped.
Flume
Select an option, then go to
Table 9
. Options: British Rectangle, Cuthroat, H-Type, Leopold-
Lagco, Montana, Palmer-Bowlus, Parshall, RBC, San Dimas or Trapezoidal.
Manning
Enter the pipe shape. Options: Circular, Rectangular, Trapezoidal or U-Shaped.
Enter the pipe diameter or the width(s) and height, percent slope and manning roughness
coefficient. Refer to
Manning roughness coefficients
on page 30.
Percent slope
—1 unit per hundred units = 0.01 slope. For example, 1 m of decline for every
100 m = 0.01 slope
Nozzle
Enter the nozzle shape.
California Pipe
—Enter the nozzle diameter (0.01 to 0.91 m).
Kennison
—Enter the nozzle diameter (6, 8, 10, 12, 16. 20, 24, 30 or 36 in.) and half-section or no half-
section.
Weir
Select an option, then go to
Table 10
. Options: Cipolletti, Compound Notch, Contracted
Rectangular, Non-Contracted Rectangle or Thel-Mar. V-Notch.
Table 8 Area velocity options
Option
Description
Circular
Enter the pipe diameter. For the US900x, in addition enter the source of the velocity measurement.
For example, when an ultrasonic sensor is connected to Sensor 1 port and a submerged area
velocity sensor is connected to Sensor 2 port, select Sensor 2 port.
Rectangular Enter the width and the height of the channel. For the US900x, in addition enter the source of the
velocity measurement.
Trapezoidal Enter the width of the top and bottom of the channel and the height of the channel. For the
US900x, in addition enter the source of the velocity measurement.
U-shaped
Enter the channel width. Options: 10 to 2540 cm (4 to 999.99 in.). For the US900x, in addition enter
the channel height and the source of the velocity measurement.
Table 9 Flume options
Option
Description
British Rectangle Enter the flume throat length, width and flume throat width.
Cutthroat
Enter the flume width and length.
H-Large
Enter the flume size. Options: 3.0 or 4.0 ft.
H-Medium
Enter the flume size. Options: 0.50, 0.75, 1.00, 1.50, 2.00, 2.50, 3.00 or 4.50 ft.
H-Small
Enter the flume size. Options: 0.4, 0.6, 0.8 or 1.0 ft.
Leopold-Lagco
Enter the flume size. Options: 4, 6, 8, 10, 12, 15, 18, 20, 21, 24, 27, 30, 36, 42, 48, 54 or
60 inches.
Montana
Enter the flume throat width. Options: 1, 2, 3, 6, 9, 12, 18, 24, 30, 36, 48, 60, 72, 84, 96, 108,
120, 144, 180, 240, 300, 360, 480 or 600 inches.
Palmer-Bowlus
Enter the flume throat size. Options: 4, 6, 8, 10, 12, 15, 18, 21, 24, 27, 30, 36, 42, 48, 60 or
72 inches.
Parshall
Enter the flume size. Options: 1, 2, 3, 6, 9, 12, 18, 24, 30, 36, 48, 60, 72, 84, 96, 108, 120,
144, 180, 240, 300, 360, 480 or 600 inches.
RBC
Enter the flume size. Options: 50, 75, 100, 150 or 200 mm.
English
9
Summary of Contents for AS950 AWRS
Page 1: ...DOC346 53 80493 AS950 Controller 06 2016 Edition 8 Operations...
Page 2: ......
Page 35: ......