PAT700 Total Organic Carbon Analyzer – Bottle Mode
Page 83
Anatel
Operator Manual
Figure 6-9 Test running dialog box
•
After verifying the presence of the appropriate standards, the analyzer performs a
temperature stabilization. After the temperature has stabilized, the blank and each
standard is analyzed in the proper sequence.
•
Once all the bottles have been analyzed, the analyzer compares the current results
to the factory calibration. If the change in slope is greater than
±
15%, the results are
rejected.
•
The dialog box displays the current status of the test and lists the results of each
repetition and bottle as it is running.
•
Cancel icon
allows you to cancel the test while it is in progress. If you cancel the
test, the analyzer turns off the pump and UV lamp, closes all valves, and notifies
you that the test has been cancelled.
11)
When the calibration is completed, the test summary dialog box appears, with the
results of the calibration and a notification of whether the calibration failed or
passed. The results also appear on the printout if a printer is attached. See
.
Figure 6-10 Test summary dialog box