Parameter setup command
Example:
Read command:
“SY”
Response:
“
7
25
2007
20
33
31
”
Write format:
“
SY
mm
DD
YYYY
HH
MM
SS
”
“
SY
7
30
2007
12
30
25
”
1.
Flush time (HH:mm:SS)
2.
Cycle time (HH:mm:SS)
3.
TOC alarm limit (ppb)
4.
Pump sampling (0 = Disabled, 1 = Enabled)
5.
Sensor name (1 to 13 characters)
6.
Idle mode (0 = No flow, 1 = Flow with conductivity, 2 = Flow without conductivity)
7.
Always 0
8.
Resistivity/conductivity units (0 = Mohm-cm, 1=
μ
S/cm C, 2 = Mohm-cm U, 3 =
μ
S U)
9.
Always 0
10. TOC mode print strategy (0 = Continuous, 1 = Percent change, 2 = Disabled)
11. Conductivity print strategy (0 = Timed, 1 = Percent change, 2 = Disabled)
12. Active alarm type (0 = TOC only, 2 = TOC and uncompensated conductivity)
13. Analog output type (0 for 4-20 mA, 1 for 0-20 mA)
14. Analog output on error (0 = Minimum [2mA], 1 = Last output, 2 = Maximum [22mA])
15. TOC percent change between printouts (1-99)
16. Conductivity percent change between printouts (1-99)
17. Analog zero-scale TOC (ppb)
18. Analog full-scale TOC (ppb)
19. User TOC calibration slope
20. Digital control mode (0 = Disabled, 1 = Enabled)
21. Always 0 (data and audit Log)
22. User conductivity calibration slope
“SY”
Set/get system date and time
“CA”
Disable automatic result
printouts for serial port
Disables the automatic printout of results for the serial port.
“HR”
Set/get global settings
Sets or gets global settings. Compatible with or without user
conductivity calibration slope parameter.